MicroScan Pos MIC Panel Type 34 - Product Usage: For the determination of antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic cocci and listeria monocytogenes.
Enforcement
- Recall Number
- Z-0720-2021
- Event ID
- 86817
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Beckman Coulter Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 30, 2020
- Initiation Date
- December 2, 2020
- Classification Date
- December 23, 2020
- Termination Date
- January 3, 2023
- Address
- 250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States
Description
MicroScan Pos MIC Panel Type 34 - Product Usage: For the determination of antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic cocci and listeria monocytogenes.
Due to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification.
Catalog Number: B1017-216 UDI Code: (01)15099590657550(17)210610(11)200610(10)2021-06-10 Lot Number: 2021-06-10
U.S. Nationwide distribution including in the states of AL, AZ, CA, FL, IN, KY, MA, MI, MO, NC, NH, NY, OH, OK, PA, SC, SD, TN, TX, VA, and WI. The countries of Australia, Austria, Bahamas, Bangladesh, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, India, Iran, Ireland, Israel, Italy, Japan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., United Kingdom, Uruguay, Vatican City, Vietnam.
685 panel boxes (20 panels per box) = 13,700 panels