FDA Enforcement Class II Terminated

LIFESTREAM Balloon Expandable Vascular Covered Stent . BARD PERIPHERAL VASCULAR. 6mm X 26mm X 80 Recommended Introducer 6F (2.00mm) NP 8atm (811 kPa) Recommended Guide wire .035" (0.89mm) RBP 12 atm (1216 kPa) RX Only Lot numbers begins with CMA or CMB. Product Code Stent Diameter Stent length Catheter Length LSM0800626 6mm 26 mm 80 cm LSM1350626 6 mm 26 mm 135 cm LSM0800637 6 mm 37 mm 80 cm LSM1350637 6 mm 37 mm 135 cm LSM0800726 7 mm 26 mm 80 cm LSM1350726 7 mm 26 mm 135 cm Product Usage: The LIFESTREAM Balloon Expandable Vascular Covered Stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries with reference vessel diameters between 4.5 mm and 12.0 mm, and lesions lengths up to 100mm.

Recall: Z-0718-2020 · Reported December 25, 2019

Enforcement

Recall Number
Z-0718-2020
Event ID
84173
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Bard Peripheral Vascular Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 25, 2019
Initiation Date
February 18, 2019
Classification Date
December 19, 2019
Termination Date
April 28, 2021
Address
1625 W 3rd St Bldg 1, Tempe, AZ, 85281-2438, United States

Description

LIFESTREAM Balloon Expandable Vascular Covered Stent . BARD PERIPHERAL VASCULAR. 6mm X 26mm X 80 Recommended Introducer 6F (2.00mm) NP 8atm (811 kPa) Recommended Guide wire .035" (0.89mm) RBP 12 atm (1216 kPa) RX Only Lot numbers begins with CMA or CMB. Product Code Stent Diameter Stent length Catheter Length LSM0800626 6mm 26 mm 80 cm LSM1350626 6 mm 26 mm 135 cm LSM0800637 6 mm 37 mm 80 cm LSM1350637 6 mm 37 mm 135 cm LSM0800726 7 mm 26 mm 80 cm LSM1350726 7 mm 26 mm 135 cm Product Usage: The LIFESTREAM Balloon Expandable Vascular Covered Stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries with reference vessel diameters between 4.5 mm and 12.0 mm, and lesions lengths up to 100mm.

Reason

Certain product code/lot number combinations may experience introducer sheath compatibility issues when used with a 6F introducer sheath. Using the affected balloon expandable vascular covered stents with a 6F introducer sheath could result in difficulty or the inability to insert or advance the catheter through the introducer sheath. This could potentially lead to a slight prolongation of the procedure.

Code Info

Begins with Lot Number CMA or CMB Product Code LSM0800626 Product Code LSM1350626 Product Code LSM0800637 Product Code LSM1350637 Product Code LSM0800726 Product Code LSM1350726

Distribution

127 units were distributed in Australia to 32 Consignees.

Quantity

127 Units