FDA Enforcement Class II Ongoing

Soltive SuperPulsed Laser System TFL FIBER CLEAVER, Model Number TFL-AFC

Recall: Z-0717-2024 · Reported January 24, 2024

Enforcement

Recall Number
Z-0717-2024
Event ID
93613
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Olympus Corporation of the Americas
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 24, 2024
Initiation Date
December 4, 2023
Classification Date
January 12, 2024
Address
3500 Corporate Pkwy, N/A, Center Valley, PA, 18034-8229, United States

Description

Soltive SuperPulsed Laser System TFL FIBER CLEAVER, Model Number TFL-AFC

Reason

The strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.

Code Info

Lot Number: All units sold prior to 2023

Distribution

US Nationwide. Global Distribution

Quantity

236 units