FDA Enforcement Class II Terminated

Peripherally Inserted Central Catheters (PICC), UPN H965750191, Catalog # 75-019 The Maximal Barrier Nursing Kits are Tandem Pack Convenience Kits packaged with a NMI PICC and accessories. The BioFlo PICC with ENDEXO Technology and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.

Recall: Z-0715-2017 · Reported December 14, 2016

Enforcement

Recall Number
Z-0715-2017
Event ID
74948
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Angiodynamics Inc. (Navilyst Medical Inc.)
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 14, 2016
Initiation Date
June 8, 2016
Classification Date
December 7, 2016
Termination Date
June 15, 2018
Address
10 Glens Falls Tech Park, N/A, Glens Falls, NY, 12801-3864, United States

Description

Peripherally Inserted Central Catheters (PICC), UPN H965750191, Catalog # 75-019 The Maximal Barrier Nursing Kits are Tandem Pack Convenience Kits packaged with a NMI PICC and accessories. The BioFlo PICC with ENDEXO Technology and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.

Reason

There is a potential for stylet stiffening that is not accurately represented on the labeling. The label states that product includes a stylet with an outside diameter (OD) measuring 0.014, however, the product may contain a stylet with an outside diameter (OD) measuring 0.016.

Code Info

Lot #: 4945781

Distribution

Nationwide Distribution to CA, FL, GA, HI, LA, MI, MN, MO, NV, OH & TX

Quantity

50 cartons (boxes of 5)