KWIK-STIK, Quality control kit for culture media, sold as: KWIK-STIK (single-pack) Enterococcus faecalis derived from ATCC 51299
Enforcement
- Recall Number
- Z-0714-2022
- Event ID
- 89451
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Microbiologics Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 9, 2022
- Initiation Date
- October 6, 2014
- Classification Date
- March 3, 2022
- Termination Date
- January 17, 2023
- Address
- 200 Cooper Ave N, Saint Cloud, MN, 56303-4440, United States
Description
KWIK-STIK, Quality control kit for culture media, sold as: KWIK-STIK (single-pack) Enterococcus faecalis derived from ATCC 51299
This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.
Catalog Number HL163263 OEM Private Label with Healthlink; UDI: N/A (OEM agreement with Healthlink ended in Q1 2016 before UDI was required for class 1 IVDs) Lot Number (Expiration Date): 959-74-1 (10/08/2014), 959-75-2 (04/22/2015), 959-75-3 (04/22/2015), 959-75-6 (04/22/2015), 959-75-7 (04/22/2015), 959-75-10 (04/22/2015), and 959-76-2 (07/08/2015)
Worldwide distribution - US Nationwide and the countries of Argentina, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia/Hrvatska, Ecuador, Finland, France, Germany, Guam, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Kenya, Republic of Korea, Latvia, Macedonia, Malaysia, Netherlands, Panama, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Singapore, South Africa, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Uganda, United Arab Emirates, and Vietnam.
150 units