Peripherally Inserted Central Catheters (PICC), UPN H965750141, Catalog # 75-014 The Maximal Barrier Nursing Kits are Tandem Pack Convenience Kits packaged with a NMI PICC and accessories. The BioFlo PICC with ENDEXO Technology and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring
Enforcement
- Recall Number
- Z-0714-2017
- Event ID
- 74948
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Angiodynamics Inc. (Navilyst Medical Inc.)
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 14, 2016
- Initiation Date
- June 8, 2016
- Classification Date
- December 7, 2016
- Termination Date
- June 15, 2018
- Address
- 10 Glens Falls Tech Park, N/A, Glens Falls, NY, 12801-3864, United States
Description
Peripherally Inserted Central Catheters (PICC), UPN H965750141, Catalog # 75-014 The Maximal Barrier Nursing Kits are Tandem Pack Convenience Kits packaged with a NMI PICC and accessories. The BioFlo PICC with ENDEXO Technology and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring
There is a potential for stylet stiffening that is not accurately represented on the labeling. The label states that product includes a stylet with an outside diameter (OD) measuring 0.014, however, the product may contain a stylet with an outside diameter (OD) measuring 0.016.
Lot #: 4941411
Nationwide Distribution to CA, FL, GA, HI, LA, MI, MN, MO, NV, OH & TX
48 cartons (boxes of 5)