FDA Enforcement Class II Terminated

Draeger Infinity Acute Care System Monitoring Solution

Recall: Z-0714-2014 · Reported January 22, 2014

Enforcement

Recall Number
Z-0714-2014
Event ID
67056
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Draeger Medical Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 22, 2014
Initiation Date
December 10, 2013
Classification Date
January 10, 2014
Termination Date
May 20, 2015
Address
3135 Quarry Rd, Telford, PA, 18969-1042, United States

Description

Draeger Infinity Acute Care System Monitoring Solution

Reason

After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

Code Info

with software versions VG2.0.3 and higher.

Distribution

US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.

Quantity

439