FDA Enforcement
Class II
Terminated
Draeger Infinity Acute Care System Monitoring Solution
Recall: Z-0714-2014
·
Reported January 22, 2014
Enforcement
- Recall Number
- Z-0714-2014
- Event ID
- 67056
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Draeger Medical Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 22, 2014
- Initiation Date
- December 10, 2013
- Classification Date
- January 10, 2014
- Termination Date
- May 20, 2015
- Address
- 3135 Quarry Rd, Telford, PA, 18969-1042, United States
Description
Draeger Infinity Acute Care System Monitoring Solution
Reason
After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.
Code Info
with software versions VG2.0.3 and higher.
Distribution
US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.
Quantity
439