Scoot Control R-net, Product Number P015-61. The Scoot Control is intended to support a wheelchair attendant in controlling or maneuvering a power chair
Recall: Z-0713-2024
·
Reported January 17, 2024
Enforcement
- Recall Number
- Z-0713-2024
- Event ID
- 93556
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- mo-Vis BVBA
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 17, 2024
- Initiation Date
- November 28, 2023
- Classification Date
- January 11, 2024
- Address
- Biebuyckstraat 15D, Nevele, N/A, Belgium
Description
Scoot Control R-net, Product Number P015-61. The Scoot Control is intended to support a wheelchair attendant in controlling or maneuvering a power chair
Reason
In rare cases, the lever of the throttle can get temporarily stuck during the movement and does not come back to its neutral point by itself. If this happens during usage, an unwanted movement of the chair could be the result. An unwanted movement could result in damage to the wheelchair, material damage around the chair and/or injury to the user, attendant or by-standers.
Code Info
UDI-DI: 05407008320119; Lot Numbers: 3304-3317 3322-3329 3440-3452 3569-3618 3830-3844 4020-4054 4170-4176 4250-4255 4321-4330
Distribution
US States: TN, TX.
Quantity
129 units