FDA Enforcement Class II Ongoing

SafeDAY set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADM SET 15 DR W/2 SAFEDAY & STOPCOCK; Catalog Number: 352646.

Recall: Z-0712-2026 · Reported December 3, 2025

Enforcement

Recall Number
Z-0712-2026
Event ID
97816
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
B Braun Medical Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 3, 2025
Initiation Date
October 29, 2025
Classification Date
November 26, 2025
Address
824 12th Ave, N/A, Bethlehem, PA, 18018-3524, United States

Description

SafeDAY set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADM SET 15 DR W/2 SAFEDAY & STOPCOCK; Catalog Number: 352646.

Reason

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Code Info

1. Catalog Number: 352646; Primary UDI-DI: 04046955626946; Unit of Dose UDI-DI: 04046955626939; Expiration Date: 36mths from Start Date.

Distribution

Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.

Quantity

554,015 units