FDA Enforcement Class II Terminated

DePuy Spine VIPER POLY SCREW 5.5 9 X100MM Tl Product Code: 1867-15-999 Spinal Implant Component.

Recall: Z-0710-2013 · Reported January 30, 2013

Enforcement

Recall Number
Z-0710-2013
Event ID
63665
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
DePuy Spine, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 30, 2013
Initiation Date
November 7, 2012
Classification Date
January 23, 2013
Termination Date
December 24, 2013
Address
325 Paramount Dr, N/A, Raynham, MA, 02767-5199, United States

Description

DePuy Spine VIPER POLY SCREW 5.5 9 X100MM Tl Product Code: 1867-15-999 Spinal Implant Component.

Reason

Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head

Code Info

All lots

Distribution

Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.

Quantity

23 units