FDA Enforcement Class II Completed

Elekta Unity

Recall: Z-0706-2019 · Reported January 16, 2019

Enforcement

Recall Number
Z-0706-2019
Event ID
81886
Classification
Class II
Status
Completed
Product Type
Devices
Firm
Elekta Limited
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
January 16, 2019
Initiation Date
January 4, 2019
Classification Date
January 9, 2019
Address
Linac House, Fleming Way, Crawley, N/A, United Kingdom

Description

Elekta Unity

Reason

In rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings during the lifetime of the system, the speakers can become dislodged and be pulled into the bore.

Code Info

Serial Numbers: 1536549, 1536549

Distribution

TX and WI, Germany, Denmark, Netherlands, United Kingdom

Quantity