FDA Enforcement
Class II
Completed
Elekta Unity
Recall: Z-0706-2019
·
Reported January 16, 2019
Enforcement
- Recall Number
- Z-0706-2019
- Event ID
- 81886
- Classification
- Class II
- Status
- Completed
- Product Type
- Devices
- Firm
- Elekta Limited
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- January 16, 2019
- Initiation Date
- January 4, 2019
- Classification Date
- January 9, 2019
- Address
- Linac House, Fleming Way, Crawley, N/A, United Kingdom
Description
Elekta Unity
Reason
In rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings during the lifetime of the system, the speakers can become dislodged and be pulled into the bore.
Code Info
Serial Numbers: 1536549, 1536549
Distribution
TX and WI, Germany, Denmark, Netherlands, United Kingdom
Quantity
2 units