FDA Enforcement
Class II
Terminated
DePuy Spine VIPER POLY SCREW 5.5 8 X70MM Tl Product Code: 1867-15-870 Spinal Implant Component.
Recall: Z-0705-2013
·
Reported January 30, 2013
Enforcement
- Recall Number
- Z-0705-2013
- Event ID
- 63665
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- DePuy Spine, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 30, 2013
- Initiation Date
- November 7, 2012
- Classification Date
- January 23, 2013
- Termination Date
- December 24, 2013
- Address
- 325 Paramount Dr, N/A, Raynham, MA, 02767-5199, United States
Description
DePuy Spine VIPER POLY SCREW 5.5 8 X70MM Tl Product Code: 1867-15-870 Spinal Implant Component.
Reason
Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head
Code Info
All lots
Distribution
Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.
Quantity
34 units