FDA Enforcement Class II Terminated

QC Sets and Panels: KWIK-STIK. Cepheid Xpert GBS QC Set (Live Culture). Sets include: 1. 0864E3 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 2. 0864E5 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 3. 0864E7 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 4. Negative Control 0243 Lactobacillus acidophilus derived from ATCC 4356

Recall: Z-0703-2022 · Reported March 9, 2022

Enforcement

Recall Number
Z-0703-2022
Event ID
89517
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Microbiologics Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 9, 2022
Initiation Date
December 1, 2015
Classification Date
March 1, 2022
Termination Date
July 7, 2022
Address
200 Cooper Ave N, Saint Cloud, MN, 56303-4440, United States

Description

QC Sets and Panels: KWIK-STIK. Cepheid Xpert GBS QC Set (Live Culture). Sets include: 1. 0864E3 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 2. 0864E5 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 3. 0864E7 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 4. Negative Control 0243 Lactobacillus acidophilus derived from ATCC 4356

Reason

The Lactobacillus acidophilus negative control (item 0243 single KWIK-STIK) contained in the 8164 Cepheid Xpert GBS QC Set and 8165 Cepheid Smart GBS QC Set was found to test positive for Group B Streptococcus (GBS) in some cases. This causes a failure when used as a quality control for the Cepheid Xpert GBS and Cepheid Smart GBS systems. The failure of the quality control test causes a disruption for the testing laboratory which may result in repeated testing or the need for alternative test methods. This recall is applicable to the 8164 and 8165 sets only.

Code Info

Catalog Number: 8164; Lot Number (Expiration Date): 8164-45 (12/16/2015), 8164-46 (12/16/2015), 8164-47 (09/02/2016), 8164-48 (09/02/2016), and 8164-49 (10/03/2016); UDI: 70845357028623

Distribution

US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, MN, NC, NM, NY, OH, OR, PA, TN, TX, VA, VT, WA, WI, WV, WY.

Quantity

541 units