FDA Enforcement Class II Ongoing

30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger, REF: CL3011; 31" (79 cm) Appx 3.4 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Bag Hanger, REF: CL3011T; 40" (102 cm) Appx 4.9 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Drip Chamber, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3020; 30" (76 cm) Appx 6.1 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, Drop-In Red Cap, REF: CL3028; 30" (76 cm) Appx 6.1 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, 0.2 Micron Filter, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3028T; 42" (107 cm) Appx 5.2 ml, Admin Set w/20 Drop In-Line Chamber, ChemoLock w/Red Cap, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3361; Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock Additive Port, Vented Cap; ChemoLock; Spinning Spiros w/Red Cap, REF: CL3538; Oncology Kit w/31" (79 cm) 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, Spiros; ChemoLock; ChemoLock Universal Vented Vial Spike; Bag Hanger, REF CL3929; 8.5" (22 cm) Appx 1.0 ml, Ext Set w/MicroClave Clear, ChemoLock Port, Spiros w/Red Cap, Clamp, REF: CL3931; 7.5" (19 cm) Appx 1.1 ml, Bifuse Ext Set w/ChemoLock Port, ChemoClave, Spiros, 2 Clamps, REF: CL3932; 30 IN(76cm)APPX 2.1ml, EXT SET w/ChemoLock w/RED CAP, Spiros RED CAP, REF: CL3935; Oncology Kit w/ChemoLock w/ChemoClave; Spinning Spiros w/ Red Cap, REF: CL3941; Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock w/Red Cap, Vented Cap; ChemoLock Port w/Bag Spike; Spiros w/Red Cap, REF: CL3946; Oncology Kit, ChemoLock Bag Spike w/Additive Port, Vented Cap, Spiros, REF: CL3948; Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros, ChemoLock, Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960; Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros, ChemoLock, Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960T; 60" (152 cm) Appx 7.4 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Chamber, Spiros w/Red Cap, Bag Hanger, REF: CL4100; 30" (76 cm) Appx 3.6 ml, Admin Set w/20 Drop In-Line Drip Chamber, ChemoLock Additive Port, Spiros w/Red Cap, REF: CL4131; 45" (114 cm) Appx 1.7 ml, Smallbore Ext Set w/Spiros w/Red Cap, Anti-Siphon Valve, Clamp, Y-Connector w/ChemoLock Port, Rotating Luer, REF: CL4151; 31" (79 cm) Appx 3.4 ml, 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, Spiros w/Red Cap, Drop-In ChemoLock Injector, Bag Hanger, REF: CL3978

Recall: Z-0703-2021 · Reported December 30, 2020

Enforcement

Recall Number
Z-0703-2021
Event ID
86426
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
ICU Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 30, 2020
Initiation Date
August 31, 2020
Classification Date
December 23, 2020
Address
951 Calle Amanecer, N/A, San Clemente, CA, 92673-6212, United States

Description

30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger, REF: CL3011; 31" (79 cm) Appx 3.4 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Bag Hanger, REF: CL3011T; 40" (102 cm) Appx 4.9 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Drip Chamber, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3020; 30" (76 cm) Appx 6.1 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, Drop-In Red Cap, REF: CL3028; 30" (76 cm) Appx 6.1 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, 0.2 Micron Filter, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3028T; 42" (107 cm) Appx 5.2 ml, Admin Set w/20 Drop In-Line Chamber, ChemoLock w/Red Cap, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3361; Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock Additive Port, Vented Cap; ChemoLock; Spinning Spiros w/Red Cap, REF: CL3538; Oncology Kit w/31" (79 cm) 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, Spiros; ChemoLock; ChemoLock Universal Vented Vial Spike; Bag Hanger, REF CL3929; 8.5" (22 cm) Appx 1.0 ml, Ext Set w/MicroClave Clear, ChemoLock Port, Spiros w/Red Cap, Clamp, REF: CL3931; 7.5" (19 cm) Appx 1.1 ml, Bifuse Ext Set w/ChemoLock Port, ChemoClave, Spiros, 2 Clamps, REF: CL3932; 30 IN(76cm)APPX 2.1ml, EXT SET w/ChemoLock w/RED CAP, Spiros RED CAP, REF: CL3935; Oncology Kit w/ChemoLock w/ChemoClave; Spinning Spiros w/ Red Cap, REF: CL3941; Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock w/Red Cap, Vented Cap; ChemoLock Port w/Bag Spike; Spiros w/Red Cap, REF: CL3946; Oncology Kit, ChemoLock Bag Spike w/Additive Port, Vented Cap, Spiros, REF: CL3948; Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros, ChemoLock, Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960; Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros, ChemoLock, Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960T; 60" (152 cm) Appx 7.4 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Chamber, Spiros w/Red Cap, Bag Hanger, REF: CL4100; 30" (76 cm) Appx 3.6 ml, Admin Set w/20 Drop In-Line Drip Chamber, ChemoLock Additive Port, Spiros w/Red Cap, REF: CL4131; 45" (114 cm) Appx 1.7 ml, Smallbore Ext Set w/Spiros w/Red Cap, Anti-Siphon Valve, Clamp, Y-Connector w/ChemoLock Port, Rotating Luer, REF: CL4151; 31" (79 cm) Appx 3.4 ml, 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, Spiros w/Red Cap, Drop-In ChemoLock Injector, Bag Hanger, REF: CL3978

Reason

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Code Info

Lots: 4720479, 4724743, 4735699, 4775051, 4803047, 4808098, 4868079, 4868080, 4735696, 4763384, 4800367, 4866943, 4908482, 4724700, 4742639, 4758414, 4775030, 4787228, 4808060, 4848204, 4712385, 4724701, 4605106, 4733557, 4757782, 4775514, 4800370, 4724728, 4765922, 4867983, 4724703, 4749489, 4803140, 4878873, 4742641, 4758385, 4794501, 4720364, 4752843, 4799894, 4850394, 4866135, 4790865, 4735669, 4742630, 4758441, 4764898, 4787232, 4848213, 4867995, 4743901, 4765927, 4775070, 4777973, 4794503, 4606697, 4735662, 4803124, 4808069, 4867961, 4897098, 4898098, 4899830, 4573189, 4606684, 4749454, 4756268, 4803136, 4808059, 4848185, 4878880, 4726940, 4742091, 4738028, 4749474, 4560357, 4735675, 4714539, 4763384, 4800367, 4866943, 4908482, 4742641

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.

Quantity

231250