FDA Enforcement Class II Ongoing

Detector H. Pylori, Catalog Number IHP-30

Recall: Z-0702-2024 · Reported January 17, 2024

Enforcement

Recall Number
Z-0702-2024
Event ID
93646
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Immunostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
January 17, 2024
Initiation Date
August 25, 2023
Classification Date
January 10, 2024
Address
38 Industrial Way E Ste 1, Eatontown, NJ, 07724-3334, United States

Description

Detector H. Pylori, Catalog Number IHP-30

Reason

Labeling for affected lots incorrectly indicates the expiration date is 10/31/2023. However, the correct expiration date is 10/31/2022. Use of the product after its expiration may cause false positive, false negative, or invalid results.

Code Info

UDI-DI 10722355002023 Lots 0821132-2 0821132-3 0821132-4 Incorrect label Expiration: 2023-10-31 Correct Expiration: 2022-10-31

Distribution

Product distributed to CA and MO.

Quantity

10 kits