FDA Enforcement Class II Ongoing

VADER pedicle system, torque wrench, Catalog Number 42-703

Recall: Z-0701-2024 · Reported January 17, 2024

Enforcement

Recall Number
Z-0701-2024
Event ID
93577
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Icotec Ag
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
January 17, 2024
Initiation Date
November 28, 2023
Classification Date
January 10, 2024
Address
Industriestrasse 12, Altstatten Sg, N/A, Switzerland

Description

VADER pedicle system, torque wrench, Catalog Number 42-703

Reason

icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.

Code Info

220803-001, 220803-002, 220803-003

Distribution

US: AZ. MI, OH, CA, CO, TX, NY, NC, DE, PA