FDA Enforcement Class II Ongoing

5 IN (13 cm) Appx 1.5 ml, Bag Spike Adapter, Spiros, REF: 20123-01; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/ Integrated Clave Drip Chamber, Clave, Spiros, REF: CH3035; 40" (102 cm) Appx 4.5 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Hanger, REF: CH3135; 14" (36 cm) Appx 4.8 ml, Bifuse Add-On Set w/Spiros w/Red Cap, Bag Spike, 3 Clamps (Blue, 2 White), Vented Cap, Drop-In Spiros w/Red Cap, REF: CH3156; 114" Bifuse Primary Set w/2 Spiros, Bag Spike, 2 20 Drop In-line Drip Chamber, Clave, Bag Hanger, REF: CH3240; 18 IN (46 cm) Appx 2.9 ml, Add-On Set w/Spiros, Clave Spike, Clamp, Cap, REF: CH3290; 27" (69 cm) Appx 5.2 ml, Bifuse Add-On Set w/15 Drop Drip Chamber, Spiros w/Red Cap, Check Valve, Vented Cap, Bag Hanger, REF: CH3294; Secondary IV Oncology Kit, REF: CH3322; 30 IN (76 cm) Appx 1.8 ml) Ext Set w/2 Spiros, 2 Red Caps, Clamp, REF: CH3327; 40" (102 cm) Appx 4.6 ml, Admin Set, 2 Spiros w/2 Red Caps, 20 Drop In-Line Drip Chamber, 2 Bag Hangers, REF: CH3373; Oncology Kit w/Spiros, Red Cap, Bag Spike w/Clave Additive Port, REF: CH3397; 45" (114 cm) Appx 5.8 ml, Ext Set w/MicroClave, 2 Clave, Spiros w/Red Cap, 2 Clamps, REF: CH3403; Oncology Kit w/20" (51 cm) Ext Set w/MicroClave, Graduated Connector, Transfer Set w/Clave Additive Port, Spiros, Red Cap, 13mm Vial Spike w/Clave, REF: CH3507; Oncology Kit w/Clave Universal Vented Vial Spike, Spinning Spiros w/Red Cap, REF: CH3546; 22" (56 cm) Add-On 150 ml Burette Set (MicroClave, No Shut-Off), w/Spiros, 2 Clamps, Vented Cap, REF: CH3554; 4.5 IN (11 cm) Appx 0.43 ml, Ext Set, Spiros w/Red Cap, REF: CH3559; Oncology Kit w/Spinning Spiros, C-Clip, Red Cap, REF: CH3581; 5 IN (13 cm) Appx 0.57 ml, Ext Set, Anti-Siphon Valve, Spiros, REF: CH3582; Oncology Kit w/Spinning Spiros w/Red Cap, C-Clip; Priming Cap, REF: CH3589; 77" (196 cm) Appx. 9.4 ml, Admin Set w/2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: CH3639; 22" (56 cm) Appx 5.6ml, Ext Set, Trifuse w/4 Clave Clear, Spiros w/Red Cap, 1.2 Micron Filter, 4 Clamps, REF: CH3770; 6.5 IN (17 cm) Appx 0.78 ml, Ext Set w/Clave Clear, Clave 4-Way Stopcock, Spiros, Red Cap, REF: CH3771; 47" (119 cm) 50 ml Diluent Set for Channel 2 w/20 Drop In-Line Drip Chamber, 2 Spiros, REF: CH4001; 47" (119 cm) 50 ml Diluent Set For Channel 2 w/Bag Spike w/MicroClave Additive Port, Spiros, REF: CH4009; 30" (76 cm) Appx 5.9 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, REF: Z7100; 30" (76 cm) 50 ml Diluent Set for Channel 2 w/Spiros, REF: CH4006; 60 IN(152cm)APPX 0.45ml,SMALLBORE EXT,MicroCLAVE,Spiros, REF: Z7040; 6" (15 cm) Appx 0.55 ml, Smallbore Trifuse Ext Set w/3 MicroClave, 3 Clamps, Spiros, REF: Z7151

Recall: Z-0701-2021 · Reported December 30, 2020

Enforcement

Recall Number
Z-0701-2021
Event ID
86426
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
ICU Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 30, 2020
Initiation Date
August 31, 2020
Classification Date
December 23, 2020
Address
951 Calle Amanecer, N/A, San Clemente, CA, 92673-6212, United States

Description

5 IN (13 cm) Appx 1.5 ml, Bag Spike Adapter, Spiros, REF: 20123-01; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/ Integrated Clave Drip Chamber, Clave, Spiros, REF: CH3035; 40" (102 cm) Appx 4.5 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Hanger, REF: CH3135; 14" (36 cm) Appx 4.8 ml, Bifuse Add-On Set w/Spiros w/Red Cap, Bag Spike, 3 Clamps (Blue, 2 White), Vented Cap, Drop-In Spiros w/Red Cap, REF: CH3156; 114" Bifuse Primary Set w/2 Spiros, Bag Spike, 2 20 Drop In-line Drip Chamber, Clave, Bag Hanger, REF: CH3240; 18 IN (46 cm) Appx 2.9 ml, Add-On Set w/Spiros, Clave Spike, Clamp, Cap, REF: CH3290; 27" (69 cm) Appx 5.2 ml, Bifuse Add-On Set w/15 Drop Drip Chamber, Spiros w/Red Cap, Check Valve, Vented Cap, Bag Hanger, REF: CH3294; Secondary IV Oncology Kit, REF: CH3322; 30 IN (76 cm) Appx 1.8 ml) Ext Set w/2 Spiros, 2 Red Caps, Clamp, REF: CH3327; 40" (102 cm) Appx 4.6 ml, Admin Set, 2 Spiros w/2 Red Caps, 20 Drop In-Line Drip Chamber, 2 Bag Hangers, REF: CH3373; Oncology Kit w/Spiros, Red Cap, Bag Spike w/Clave Additive Port, REF: CH3397; 45" (114 cm) Appx 5.8 ml, Ext Set w/MicroClave, 2 Clave, Spiros w/Red Cap, 2 Clamps, REF: CH3403; Oncology Kit w/20" (51 cm) Ext Set w/MicroClave, Graduated Connector, Transfer Set w/Clave Additive Port, Spiros, Red Cap, 13mm Vial Spike w/Clave, REF: CH3507; Oncology Kit w/Clave Universal Vented Vial Spike, Spinning Spiros w/Red Cap, REF: CH3546; 22" (56 cm) Add-On 150 ml Burette Set (MicroClave, No Shut-Off), w/Spiros, 2 Clamps, Vented Cap, REF: CH3554; 4.5 IN (11 cm) Appx 0.43 ml, Ext Set, Spiros w/Red Cap, REF: CH3559; Oncology Kit w/Spinning Spiros, C-Clip, Red Cap, REF: CH3581; 5 IN (13 cm) Appx 0.57 ml, Ext Set, Anti-Siphon Valve, Spiros, REF: CH3582; Oncology Kit w/Spinning Spiros w/Red Cap, C-Clip; Priming Cap, REF: CH3589; 77" (196 cm) Appx. 9.4 ml, Admin Set w/2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: CH3639; 22" (56 cm) Appx 5.6ml, Ext Set, Trifuse w/4 Clave Clear, Spiros w/Red Cap, 1.2 Micron Filter, 4 Clamps, REF: CH3770; 6.5 IN (17 cm) Appx 0.78 ml, Ext Set w/Clave Clear, Clave 4-Way Stopcock, Spiros, Red Cap, REF: CH3771; 47" (119 cm) 50 ml Diluent Set for Channel 2 w/20 Drop In-Line Drip Chamber, 2 Spiros, REF: CH4001; 47" (119 cm) 50 ml Diluent Set For Channel 2 w/Bag Spike w/MicroClave Additive Port, Spiros, REF: CH4009; 30" (76 cm) Appx 5.9 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, REF: Z7100; 30" (76 cm) 50 ml Diluent Set for Channel 2 w/Spiros, REF: CH4006; 60 IN(152cm)APPX 0.45ml,SMALLBORE EXT,MicroCLAVE,Spiros, REF: Z7040; 6" (15 cm) Appx 0.55 ml, Smallbore Trifuse Ext Set w/3 MicroClave, 3 Clamps, Spiros, REF: Z7151

Reason

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Code Info

Lots: 4726847 4764618 4749803 4882578 4719415 4858408 4763254 4799746 4909239 4763263 4807851 4849307 4711357 4734298 4784307 4877040 4603746 4751052 4809068 4719780 4887803 4902877 4741245 4773197 4793534 4853586 4603189 4749895 4800302 4749463 4751098 4775532 4853625 4877133 4902884 4726180 4751919 4763325 4742008 4763328 4877406 4902270 4589996 4741273 4751930 4764078 4849398 4894536 4742017 4784331 4894544 4774127 4719576 4787217 4908406 4752847 4866868 4908415 4854840 4853711 4894570 4759148 4605923 4784875 4800246 4807647 4866700 4590960 4802679 4889077

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.

Quantity

55000