FDA Enforcement Class II Terminated

3M Universal Electrosurgical Pad, REF 9130 (100 pads / case) and 3M Universal Electrosurgical Pad with Cord, REF 9135-LP (40 pads / case) Product Usage: 3M Universal Electrosurgical Pads are designed to work with most electrosurgical unite (ESU) for virtually every surgical application where electrosurgery is utilized to provide a safe return path for electrosurgical current. Solid Universal Electrosurgical Pads are for use with generators that do not have a Contact Quality Monitoring System (CQMS). Split Style Universal Electrosurgical Pad is for use with ESUs that have a CQMS (i.e. REM, ARM, NESSY, etc.). 3M Universal Electrosurgical Pads are designed to be used on any patient where full skin contact and a suitable placement site can be obtained. There are no patient weight restrictions for use of this product. Use of this product for unintended application could lead to an unsafe condition.

Recall: Z-0701-2017 · Reported December 14, 2016

Enforcement

Recall Number
Z-0701-2017
Event ID
75644
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
3M Company - Health Care Business
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
December 14, 2016
Initiation Date
October 26, 2016
Classification Date
December 3, 2016
Termination Date
August 1, 2017
Address
3m Center 2510 Conway Ave , B# 275-5-W-6, N/A, Saint Paul, MN, 55144-0001, United States

Description

3M Universal Electrosurgical Pad, REF 9130 (100 pads / case) and 3M Universal Electrosurgical Pad with Cord, REF 9135-LP (40 pads / case) Product Usage: 3M Universal Electrosurgical Pads are designed to work with most electrosurgical unite (ESU) for virtually every surgical application where electrosurgery is utilized to provide a safe return path for electrosurgical current. Solid Universal Electrosurgical Pads are for use with generators that do not have a Contact Quality Monitoring System (CQMS). Split Style Universal Electrosurgical Pad is for use with ESUs that have a CQMS (i.e. REM, ARM, NESSY, etc.). 3M Universal Electrosurgical Pads are designed to be used on any patient where full skin contact and a suitable placement site can be obtained. There are no patient weight restrictions for use of this product. Use of this product for unintended application could lead to an unsafe condition.

Reason

3M is recalling Universal Electrosurgical Pads because a report of an unintended material (process liner) in the product. This may prevent or impede the safe return of electrosurgical current following Electrosurgical Unit (ESU) activation. This defect has the potential to increase the risk of a patient burn directly under the pad or at an alternate site of the body.

Code Info

Product REF 9130 Lot Numbers: 2019-02DT (Expiration 02-2019) and 2019-03DU (Expiration 03-2019) Product REF 9735-LP Lot Numbers: 2019-02DB and 2019-02CY. (Both with Expiration 02-2019)

Distribution

Worldwide Distribution - US Nationwide in the states of AR, AZ, CA, FL, GA, ID, IL, KY, NC, NY, TN, TX, VA and the countries of Canada, Chile, Costa Rica, Germany, Japan, Mexico and South Africa.

Quantity

32,000 pads