FDA Enforcement Class III Ongoing

Blazer Open-Irrigated Ablation Batch Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system

Recall: Z-0699-2019 · Reported January 16, 2019

Enforcement

Recall Number
Z-0699-2019
Event ID
81747
Classification
Class III
Status
Ongoing
Product Type
Devices
Firm
Boston Scientific Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 16, 2019
Initiation Date
November 28, 2018
Classification Date
January 8, 2019
Address
300 Boston Scientific Way, Marlborough, MA, 01752-1291, United States

Description

Blazer Open-Irrigated Ablation Batch Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system

Reason

Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.

Code Info

Blazer Open Irrigated Ablation Catheter Material Number: M00496200, Batch: 20399570, Expiration Date: 3/13/2020 Batch: 21778776, Expiration Date: 2/21/2021 Batch: 22152470, Expiration Date: 5/21/2021 Batch: 22499259, Expiration Date: 8/13/2001 Material Number: M0049620N40 Batch::20413170, Expiration Date: 3/16/2020 Batch: 20809676, Expiration Date: 6/22/2020 Material Number: M0049620K20 Batch: 20403945, Expiration Date: 3/13/2020

Distribution

Japan

Quantity

1,282 (all three types in Japan)