Microbiologics QC Sets and Panels. BD MAX" CT/GC/TV Verification and Control Panel. Containing High, Medium and Low Level Controls.
Recall: Z-0698-2022
·
Reported March 9, 2022
Enforcement
- Recall Number
- Z-0698-2022
- Event ID
- 89446
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Microbiologics Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 9, 2022
- Initiation Date
- June 6, 2017
- Classification Date
- March 1, 2022
- Termination Date
- January 17, 2023
- Address
- 200 Cooper Ave N, Saint Cloud, MN, 56303-4440, United States
Description
Microbiologics QC Sets and Panels. BD MAX" CT/GC/TV Verification and Control Panel. Containing High, Medium and Low Level Controls.
Reason
Real-time shelf life testing failed at 24 months
Code Info
Catalog Number: 8193; Lot Number: 8193-04, 8193-06; UDI: 70845357041080
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, CA, DC, FL, IA, ID, KS, LA, MD, MN, MO, NE, NH, NJ, NY, OH, OK, PA, SC, TX, VA, WA, WV and the countries of Australia, Belgium, Canada, India, United Kingdom.
Quantity
29 units