FDA Enforcement Class III Terminated

Roche COBAS INTEGRA 400 and 400 plus Analyzer, an in vitro diagnostic analyzer.

Recall: Z-0696-2013 · Reported January 30, 2013

Enforcement

Recall Number
Z-0696-2013
Event ID
63617
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Roche Diagnostics Operations, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 30, 2013
Initiation Date
November 19, 2012
Classification Date
January 18, 2013
Termination Date
January 13, 2014
Address
9115 Hague Road, N/A, Indianapolis, IN, 46256-1025, United States

Description

Roche COBAS INTEGRA 400 and 400 plus Analyzer, an in vitro diagnostic analyzer.

Reason

A software security issue with Oracles TNS-Listener component has been identified. TNS-Listener is supplied with the Roche COBAS INTEGRA 400/400 plus Analyzers. The Oracle TNS-Listener can be considered as an authentication and redirection component between the client and the database inside the Roche product. This means an attacker could register their own non-Roche database via remote access

Code Info

Part Numbers: 28065047001 and 03245233001

Distribution

Nationwide Distribution including AL, AR, AZ, CA, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR.

Quantity

845 Units