FDA Enforcement Class II Terminated

D-10 Rack Loader; Model Number 220-0600; Hematology: The D-10 is a fully integrated system for sample dilution, processing and analysis of hemoglobin, designed for use with specific Bio-Rad reagent kits. The D-10 incorporates the use of a dedicated software system for instrument control, data collection, and analysis.

Recall: Z-0695-2016 · Reported February 3, 2016

Enforcement

Recall Number
Z-0695-2016
Event ID
72927
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Bio-Rad Laboratories, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
February 3, 2016
Initiation Date
December 15, 2015
Classification Date
January 27, 2016
Termination Date
May 2, 2016
Address
4000 Alfred Nobel Dr, N/A, Hercules, CA, 94547-1803, United States

Description

D-10 Rack Loader; Model Number 220-0600; Hematology: The D-10 is a fully integrated system for sample dilution, processing and analysis of hemoglobin, designed for use with specific Bio-Rad reagent kits. The D-10 incorporates the use of a dedicated software system for instrument control, data collection, and analysis.

Reason

On a rare occasion, there is a potential to assign a patient result to an incorrect sample ID when running in D10 Rack Loader configuration.

Code Info

All codes - 2006 to present

Distribution

Worldwide Distribution-US (nationwide) and the countries of Canada, Australia, Belgium, Brazil, Brunei Darussalam, China, Czech Republic, France, Germany, Greece, India, Israel, Italy, Malaysia, Maldives, New Zealand, Norway, Austria, Poland, Portugal, Russia, Spain, Sweden, Switzerland, Thailand, and UK.

Quantity

165 in US, 200 Internationally