FDA Enforcement Class II Terminated

SOMATOM Force, System x-ray, tomography, computed

Recall: Z-0692-2017 · Reported December 14, 2016

Enforcement

Recall Number
Z-0692-2017
Event ID
75594
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 14, 2016
Initiation Date
October 20, 2016
Classification Date
December 2, 2016
Termination Date
May 30, 2017
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States

Description

SOMATOM Force, System x-ray, tomography, computed

Reason

Siemens is providing software update version VA50A_SP3 to address the software bugs that were identified through normal field monitoring and the Global Complaint Handling Process. Correction for the problems are as follows: 1. Correction to volumetric misrepresentations of high contrast objects when using ADMIRE. 2. Correction to highly sporadic scan aborts due to temporarily tube currents at 0mA. 3. Correction to missing Microsoft Hotfixes (MS16-001, MS15-135, MS15-088, MS15-048). 4. Correction to highly sporadic image artifacts caused by error signals from high voltage chain at extremely high tube currents. 5. Correction to highly sporadic incomplete image reconstruction due to temporary slow hard disk write speed. 6. Correction to highly sporadic scan aborts due to temporarily high disk load of system disk. 7. Correction to highly sporadic image artifacts generated by scatter correction in case of tube arcings.

Code Info

Model Number 10742326, Serial Numbers: 75708,75709,75607,75617,75550,75526 75512,75567,75536,75711,75514,75600,75563,75510,75454,75540,75571,75500 75527,75475,75535,75481,75482,75710,75460,75476,75588,75493,75458,75528 75467,75602,75487,75464,75478,75604,75513,75532,75555,75524,75515,75559 75610,75501,75570,75616,75576,75605

Distribution

Nationwide Distribution.

Quantity

48 units distributed in U.S.