FDA Enforcement
Class I
Terminated
Emergency Transport Ventilator Oxylog 2000 Plus Ventilator Catalog Number: 5705080. Volume controlled emergency and transport ventilator with pressure support.
Recall: Z-0691-2016
·
Reported February 10, 2016
Enforcement
- Recall Number
- Z-0691-2016
- Event ID
- 72914
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Draeger Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 10, 2016
- Initiation Date
- December 22, 2015
- Classification Date
- February 3, 2016
- Termination Date
- April 5, 2017
- Address
- 3135 Quarry Rd, N/A, Telford, PA, 18969-1042, United States
Description
Emergency Transport Ventilator Oxylog 2000 Plus Ventilator Catalog Number: 5705080. Volume controlled emergency and transport ventilator with pressure support.
Reason
Drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.
Code Info
All Oxylog 2000 plus Ventilators with the identified catalog numbers distributed between April 2007 - December 2015.
Distribution
Nationwide Distribution -- including states of AR, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
Quantity
0