FDA Enforcement Class I Terminated

Emergency Transport Ventilator Oxylog 2000 Plus Ventilator Catalog Number: 5705080. Volume controlled emergency and transport ventilator with pressure support.

Recall: Z-0691-2016 · Reported February 10, 2016

Enforcement

Recall Number
Z-0691-2016
Event ID
72914
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Draeger Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 10, 2016
Initiation Date
December 22, 2015
Classification Date
February 3, 2016
Termination Date
April 5, 2017
Address
3135 Quarry Rd, N/A, Telford, PA, 18969-1042, United States

Description

Emergency Transport Ventilator Oxylog 2000 Plus Ventilator Catalog Number: 5705080. Volume controlled emergency and transport ventilator with pressure support.

Reason

Drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.

Code Info

All Oxylog 2000 plus Ventilators with the identified catalog numbers distributed between April 2007 - December 2015.

Distribution

Nationwide Distribution -- including states of AR, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY.

Quantity

0