FDA Enforcement Class II Terminated

Titanium MPJ (Toe Implant) TOE JOINT TITANIUM PC M/L, Part ID 17037

Recall: Z-0688-2022 · Reported March 9, 2022

Enforcement

Recall Number
Z-0688-2022
Event ID
89605
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
BioPro, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 9, 2022
Initiation Date
February 1, 2022
Classification Date
February 28, 2022
Termination Date
February 28, 2024
Address
2929 Lapeer Rd, Port Huron, MI, 48060-2558, United States

Description

Titanium MPJ (Toe Implant) TOE JOINT TITANIUM PC M/L, Part ID 17037

Reason

There is an incorrect statement regarding MR Compatibility in the BioPro Titanium MPJ implant IFU (IFU BP102303 rev 05) and patient implant cards (22266 rev 00) which indicates that the BioPro MPJ Implant has been tested for safety and compatibility in the MR Environment and is MR Conditional. MR Testing has not been conducted on the titanium version of the MPJ.

Code Info

Lot Number: 124737, 125470; UDI: M20917037

Distribution

US Nationwide distribution in the states of CT, KY, NM, PA, TN, and SD.