FDA Enforcement
Class II
Terminated
DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct.
Recall: Z-0685-2013
·
Reported January 23, 2013
Enforcement
- Recall Number
- Z-0685-2013
- Event ID
- 62719
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- DePuy Spine, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 23, 2013
- Initiation Date
- July 24, 2012
- Classification Date
- January 16, 2013
- Termination Date
- December 30, 2013
- Address
- 325 Paramount Dr, N/A, Raynham, MA, 02767-5199, United States
Description
DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct.
Reason
A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification.
Code Info
All lots
Distribution
Worldwide Distribution - USA (nationwide) and internationally to Belgium, Egypt, Ireland, Italy, Portugal, and the UK.
Quantity
140 units