Walgreens At-Home A1C Test Kit, Software Version Revision D, REF Number 3056
Enforcement
- Recall Number
- Z-0684-2020
- Event ID
- 84231
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Polymer Technology Systems
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 25, 2019
- Initiation Date
- December 2, 2016
- Classification Date
- December 13, 2019
- Termination Date
- July 16, 2020
- Address
- 4600 Anson Blvd, Whitestown, IN, 46075, United States
Description
Walgreens At-Home A1C Test Kit, Software Version Revision D, REF Number 3056
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
Lot Numbers: 1606405 1610622 1610910
The products were distributed to the following US states: AR, CA, FL, IL, IN, KY, MN, NC, NJ, NY, OH, PA, RI TN, UT, VA, WA, and Puerto Rico. The products were distributed to the following foreign countries: Australia, Bahamas, Belgium, Bahrain, Bangladesh, Canada, China, Columbia, Czech Republic, England, Egypt, Greece, Spain, India, Indonesia, Haiti, Hong Kong, Italy, Japan, Jordan, Kenya, Lebanon, Mexico, Myanmar, Malaysia, Philippines, Pakistan, South Africa, Nigeria, Poland, Saudi Arabia, Sweden, Turkey, Singapore, Trinidad & Tobago, Slovenia, Vietnam, United Arab Emirates, Uruguay.
15,118