FDA Enforcement Class III Terminated

Dynasty Liners: REF DLCOGA28, REF DLCOGB32, REF DLCOGC36, REF DLCOGD38, REF DLCOGE40, REF DLCOGF42, REF DLCOGG44, REF DLCOGH48, REF DLCOGJ52, REF DLCOGK56, For use with the conserve metal head

Recall: Z-0684-2017 · Reported December 7, 2016

Enforcement

Recall Number
Z-0684-2017
Event ID
75063
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
MicroPort Orthopedics Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 7, 2016
Initiation Date
June 21, 2016
Classification Date
December 1, 2016
Termination Date
July 1, 2017
Address
5677 Airline Rd, Arlington, TN, 38002-9501, United States

Description

Dynasty Liners: REF DLCOGA28, REF DLCOGB32, REF DLCOGC36, REF DLCOGD38, REF DLCOGE40, REF DLCOGF42, REF DLCOGG44, REF DLCOGH48, REF DLCOGJ52, REF DLCOGK56, For use with the conserve metal head

Reason

Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.

Code Info

Serial Numbers: REF DLCOGA28 Seral Numbers: 1509498, 1509897, 107461555, 117476083, 127477625, 0411341968, 0611341969, 0901199714 REF DLCOGB32 Serial Numbers: 1428321, 027403682, 027403684, 027412214, 027412216, 027412218, 027412220, 027412222, 029789268, 048571448, 058605178, 068625411, 069837383, 078628910, 079859721, 087467577, 097474918, 097476396, 097477626, 098657540, 099905615, 099927666, 0101018635, 107480309, 107486655, 109968495, 0111248666, 0111254637, 117486656, 117501983, 119982392, 127504540, 127508429, 0211298850, 0211298851, 0401072608, 0501123237, 0601150684, 0711370347, 0711370832, 0711371855, 0711388872, 0801184970, 0801185465, 0901196484, 0901199209, 0901200444, 0901200704 REF DLCOGC36 Serial Numbers: 1389336, 1411596, 1412300, 1419309, 1421706, 1421872, 1421873, 1425040, 1425041, 1426635, 1441425, 1447128, 1448009, 017395680, 017396168, 017398843, 019752394, 027406078, 027406080, 027406549, 027408437, 027408506, 027412667, 029797073,038565179, 038571449, 039808007, 039823319, 039830311, 058576611, 059835277, 059843767, 059859139, 059869515, 067444752, 068608657, 069876613, 069885333, 077449136, 078628911, 078648401, 078648402, 086356142, 086364960, 087462096, 089905617, 089920491, 089924903, 097475888, 097475891, 097475892, 099932177, 099937595, 0101009548, 108705038, 109966090, 0111273375, 0111274786, 118728271, 119975127, 127506074, 127508430, 127514978, 129979634, 129990530, 0201044377, 0201044378, 0201048252, 0201071222, 0211301143, 0301068674, 0301084452, 0401085769, 0501125124, 0611343376, 0611343377, 0611365354, 0611366004, 0611366646, 0701164679, 0701165562, 0701178783, 0801190509, 0901189893, 0901191386, 0901200880, 0901201441, 0901201663 REF DLCOGD38 Serial Numbers: 1390011, 1422255, 1422256, 017395681, 017396169, 019751147, 019752153, 027404999, 027405000, 027408507, 027408521, 027411898, 027412100, 029793006, 038565180, 038571452, 039808008, 039823320, 039831045, 048576612, 049839752, 059846062, 059846677, 059868799, 059869517, 067445955, 067447299, 068608658, 069884531, 069888788, 077459273, 078628912, 078648403, 087464200, 087466701, 089901144, 089911489, 089922646, 096362335, 096374885, 097472362, 097474113, 097476397, 099932178, 099933792, 099937596, 0101038419, 0101038420, 107482012, 107486663, 108706223, 108716643, 108720178, 109941938, 109970863, 0111274868, 0111274869, 117486664, 117489826, 117501984, 118731716, 119980968, 127466702, 127504541, 127508431, 127512298, 128748672, 129990531, 0201062908, 0301062907, 0301067280, 0301067281, 0301081677, 0401096840, 0401098166, 0401109395, 0501127076, 0511344345, 0601147758, 0601148281, 0601148932, 0611344346, 0611346090, 0611346455, 0701172549, 0801192624, 0801193584, 0801193987, 0801194817, 0901192045, 0901203167, 0901206932, 1001206425, 1291016257, 1291016258, 1291019953, 1291019954, 08991148910 REF DLCOGE40 Serial Numbers: 1383898, 1422893, 017395849, 017396101, 017397513, 017398320, 019747416, 019749028, 019752393, 027404377, 027404964, 027406550, 027406551, 027407780, 039802355, 039809788, 039826200, 048571453, 048571454, 049834730, 049839160, 067436735, 069859138, 069868803, 069869518, 076357782, 077440680, 077453681, 078595478, 078595479, 078648404, 078648405, 079888789, 086362336, 087460119, 087460601, 089904051, 089920496, 096372317, 097472371, 097476398, 099931104, 099933793, 0101022064, 0101022065, 0101030547, 0101030548, 0101042594, 0101042595, 107482014, 107486667, 107486668, 107489825, 109941939, 109944905, 0111271873, 0111272250, 117489816, 117501985, 118722756, 118731717, 119980969, 127504542, 127508432, 127512297, 129990532, 0201057245, 0201064292, 0211295490, 0211301165, 0211302281, 0301064291, 0301081699, 0301083868, 0301084453, 0301095292, 0401095291, 0401096841, 0601165833, 0601165834, 0611361690, 0611361691, 0611364735, 0701170121, 0701170775, 0701171294, 0801194833, 0901194835, 0901207002, 0901208793, 0901208953 REF DLCOGF42 Serial Numbers: 1199809, 1366660, 1422894, 1423547, 017396170, 017397514, 017397515, 017398844, 027406424, 027406553, 027411519, 027411895, 027411896, 027413499, 029789333, 029789334, 029798203, 029809789, 029814314, 029819342, 037413500, 039819343, 049833519, 049839753, 058608659, 059846676, 059857910, 068595480, 069883241, 077449093, 078634682, 078634683, 078648406, 079860580, 079887540, 079894027, 086370088, 087462468, 089914628, 096372318, 097472373, 097476399, 099904052, 099931103, 0101022066, 0101030549, 0101030550, 107482016, 107486669, 107486670, 108705039, 108716644, 109922647, 109937597, 109943610, 109966091, 0111276618, 117496631, 117501986, 118728903, 118744506, 119972083, 127504543, 127509428, 127514975, 127514976, 128749026, 128751278, 129990533, 0201053787, 0201053788, 0201062909, 0201062910, 0201068675, 0211299901, 0301071223, 0301081700, 0301083869, 0401104095, 0401104096, 0501127409, 0611346456, 0701167385, 0701168088, 0701168230, 0701169430, 0701170051, 0711367085, 0901208997, 0901209918, 0901209977, 1101247355, 1101248443, 010103054910 REF DLCOGG44 Serial Numbers: 1383128, 1383366, 1392621, 1428789, 017396171, 017397398, 027404375, 027404965, 027406552, 027408531, 027412079, 027412080, 027412086, 029792981, 029793005, 029799054, 029799055, 038565181, 039828320, 039832455, 048571456, 048576613, 049828319, 049830312, 049839159, 057438911,058608660, 059855232, 059857911, 059860406, 059869519, 067445525, 069888790, 077447460, 077459517, 078608662, 078628913, 078648407, 078657546, 079894757, 087464126, 089915880, 089922080, 096362338, 096374886, 097476400, 097476401, 097480312, 098694868, 098706224, 099932179, 099934263, 099939751, 0101025139, 0101025144, 0101027410, 0101038421, 0101038422, 0101046891, 107481184, 107486653, 107487346, 108705037, 108716645, 108723278, 109966092, 117495711, 117495712, 117505591, 118730735, 118744505, 119975128, 119975129, 127504544, 127513338, 127518295, 127522391, 128747417, 128748669, 129998770, 129998777, 0201057246, 0201059716, 0201059717, 0201076758, 0211275710, 0301076757, 0301085876, 0301085877, 0301099035, 0301099036, 0311294091, 0311295489, 0401099037, 0401100077, 0401100078, 0401106374, 0401112729, 0411306527, 0411307069, 0601153393, 0601154045, 0601155591, 0801181073, 0801181695, 0801181920, 0801183924, 0801199319, 0901210490, 1001213377, 1001214138, 1101232813, 1201256582, 1201256583, 1201257832, 1291000319, 1291000321, 1291025140 REF DLCOGH48 Serial Numbers: 027403685, 027403686, 027410682, 027412223, 027412224, 027412225, 027412226, 078608663, 078648408, 078648409, 089904664, 097471988, 097476402, 097476542, 097477627, 0101042593, 107481183, 109967195, 0111277921, 117486671, 117496632, 127504545, 127508400, 127518297, 0201072609, 0401089450, 0601147117, 0901187425, 1101231165, 1291014716 REF DLCOGJ52 Serial Numbers: 099898795, 0101023111, 0201049895, 0201049896 REF DLCOGK56 Serial Numbers: 099898794, 0201023112, 0201049903, 0201052602, 0201052603

Distribution

Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.

Quantity

6,250 units