FDA Enforcement
Class II
Terminated
Titanium MPJ (Toe Implant) TOE JOINT TITANIUM NPC MD, Part ID 16814
Recall: Z-0683-2022
·
Reported March 9, 2022
Enforcement
- Recall Number
- Z-0683-2022
- Event ID
- 89605
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- BioPro, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 9, 2022
- Initiation Date
- February 1, 2022
- Classification Date
- February 28, 2022
- Termination Date
- February 28, 2024
- Address
- 2929 Lapeer Rd, Port Huron, MI, 48060-2558, United States
Description
Titanium MPJ (Toe Implant) TOE JOINT TITANIUM NPC MD, Part ID 16814
Reason
There is an incorrect statement regarding MR Compatibility in the BioPro Titanium MPJ implant IFU (IFU BP102303 rev 05) and patient implant cards (22266 rev 00) which indicates that the BioPro MPJ Implant has been tested for safety and compatibility in the MR Environment and is MR Conditional. MR Testing has not been conducted on the titanium version of the MPJ.
Code Info
Lot Number: 125423, 128356; UDI: 00810012480093
Distribution
US Nationwide distribution in the states of CT, KY, NM, PA, TN, and SD.