FDA Enforcement Class II Terminated

20Ga x .75 LifeGuard Huber Needle with Y-site. Product Code: CLGY-2034.

Recall: Z-0679-2018 · Reported February 28, 2018

Enforcement

Recall Number
Z-0679-2018
Event ID
78951
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Vygon MFG, Inc., dba/ Churchill Medical Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
February 28, 2018
Initiation Date
July 24, 2017
Classification Date
February 21, 2018
Termination Date
November 7, 2019
Address
87 Venture Dr, N/A, Dover, NH, 03820-5914, United States

Description

20Ga x .75 LifeGuard Huber Needle with Y-site. Product Code: CLGY-2034.

Reason

Vygon is conducting a voluntary recall of the products and lots listed in this recall, being initiated because our Tyvek supplier has notified Vygon of issues with their product, specifically the Tyvek may not be coated completely causing gaps in the adhesive used to seal the product packaging. This potential lack of adhesive may affect the sterility of the product.

Code Info

Lot Number: 1706017D

Distribution

U.S.

Quantity

396 units