FDA Enforcement Class II Terminated

Computed Tomography X-Ray System Brilliance Workspace Portal, Model #728269 Philips Medical Systems, Highland Heights, OH. The Brilliance Workspace Portal is a component of the Computed Tomography X-Ray System that is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. The portal (together with the scanner) provides image processing and display through software applications that process, analyze, display, quantify and interpret medical images/data.

Recall: Z-0675-2013 · Reported January 23, 2013

Enforcement

Recall Number
Z-0675-2013
Event ID
63393
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 23, 2013
Initiation Date
September 11, 2012
Classification Date
January 15, 2013
Termination Date
February 14, 2014
Address
595 Miner Rd, N/A, Cleveland, OH, 44143-2131, United States

Description

Computed Tomography X-Ray System Brilliance Workspace Portal, Model #728269 Philips Medical Systems, Highland Heights, OH. The Brilliance Workspace Portal is a component of the Computed Tomography X-Ray System that is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. The portal (together with the scanner) provides image processing and display through software applications that process, analyze, display, quantify and interpret medical images/data.

Reason

Philips was notified that when performing post processing with the Brilliance Workspace Portal platform with software versions 2.6.1.4, the coronary arteries in the rendered image are segmented automatically. However, the center line is not extracted or highlighted upon hovering and making it impossible to activate the vessel and choose the appropriate name (label) from the drop-down menu.

Code Info

Model #728269; Serial #s: 79906, 79921, 79927, 79937, 80007, 80026, 80031, 80040, 80058, 80062, 80066, 80069, 80087, 80110, 80121, 80155, 80167, 80168, 80169, 80174, 80192, 80193, 80195, 80201, 80208, 80209, 80219, 80227, 80228, 80230, 80252, 80258, 80266, 80280, 80289, 80293, 80294, 80315, 80322, 80328, 80333, 80340, 80341, 80348, 80360, 80375, 80385, 80387, 80394, 80398, 80416, 80421, 80422, 80429, 80447, 80451, 80460, 80474, 80482, 80484, 80485, 80492, 80497, 80505, 80516, 80578, 80585, 80591, 80629, 80646, 80658, 80681, 80704, 80709, 80711, 80720, 80723, 80743, 80749, 80751, 80831, 80835, 80866, 80878, 80908, 80909, 80911, 80914, 80916, 80930, 80931, 80944, 80974, 80992, 80999, 81023, 81025, 81034, 81045, 81071, 81099, 81101, 81107, 81117, 81126, 81136, 95382 & 95592.

Distribution

Worldwide Distribution-USA including the states of AZ, CA, CO, IL, IL, IN, MA, ME, MI, MO, NC, NJ, NY, OR, TN, TX, VA and WI and the countries of Austria, Canada, China, France, India, Ireland, Italy, Japan, Netherlands, Norway, Spain, Switzerland, Thailand and United Kingdom.

Quantity

108 units