FDA Enforcement Class II Ongoing

Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma.

Recall: Z-0673-2025 · Reported December 18, 2024

Enforcement

Recall Number
Z-0673-2025
Event ID
95738
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 18, 2024
Initiation Date
November 8, 2024
Classification Date
December 11, 2024
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232, United States

Description

Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma.

Reason

The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the intact PTH (parathyroid hormone) test may produce falsely decreased intact parathyroid hormone results. This could lead to a missed diagnosis, a delay in treatment, or inappropriate clinical management, depending upon the patient s clinical condition.

Code Info

UDI-DI 15099590201937 Lots 338921 339071 372223 439138 439205 439206 439207 439650 439801 439889 440026 440090 472012 472092 472136 472154

Distribution

Foreign distribution to Albania, Australia, Austria, Bahrain, Belgium, Brazil, C¿te d'Ivoire, Croatia, Czechia, Egypt, France, Germany, Ghana, Hungary, India, Ireland, Israel, Italy, Kazakhstan, Kuwait, Malaysia, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, S. Korea, Spain, Switzerland, Taiwan, United Kingdom of Great Britain and Northern Ireland.

Quantity

31 US; 167 OUS