Bausch and Lomb 27G sterile cannula packed in the Bausch and Lomb Amvisc 1.2% Sodium Hyaluronate ( Model 59051, 59081, 59051L, 59081L) and Amvisc Plus 1.6% Sodium Hyaluronate (Model 60081, 60051, 60051L, 60081L) ophthalmic viscosurgical device. The 27G disposable cannula is attached to the syringe standard luer fitting to deliver the Amvis or Amvis Plus during use as a surgical aid in opthalmic anterior or posterior segment procedures.
Enforcement
- Recall Number
- Z-0672-2013
- Event ID
- 63698
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Bausch and Lomb, Incorporated
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- January 23, 2013
- Initiation Date
- November 15, 2012
- Classification Date
- January 17, 2013
- Termination Date
- December 4, 2013
- Address
- 30 Enterprise Ste 450, N/A, Aliso Viejo, CA, 92656-7115, United States
Description
Bausch and Lomb 27G sterile cannula packed in the Bausch and Lomb Amvisc 1.2% Sodium Hyaluronate ( Model 59051, 59081, 59051L, 59081L) and Amvisc Plus 1.6% Sodium Hyaluronate (Model 60081, 60051, 60051L, 60081L) ophthalmic viscosurgical device. The 27G disposable cannula is attached to the syringe standard luer fitting to deliver the Amvis or Amvis Plus during use as a surgical aid in opthalmic anterior or posterior segment procedures.
Some disposable cannulas (with 4-digit lot numbers) provided with the Amvisc or Amvisc Plus OVD are not properly engaging to the Luer-Lock on the Amvisc and Amvisc Plus sterile glass syringes. The cannulas may leak viscoelastic material or detach from the syringe during injection. In rare incidents, detachment has resulted in serious patient injury.
Lot Numbers: 23632, 23633, 23639, 23640, 23647, 23801, 23811, 23863, 23886, 23897, 23920, 23933, 23972, 24022, 24025, 24028, 24257, 24323, 24436, B100003C, B100003E, B100003G, B100015A, B100015D, B100023A, B100023B, B100023C, B100023D, B100023E, B100039C, B100039D, B100039F, B100046A, B100046B, B100046C, B100058A, B100069A, B100069E, B110027D, B110028C, B110048A, B110048D, B110056A, B120011D, B120016A, B120046C, C100039A, C100039B, C10003B, C100046A, C100046B, C100046C, D090095B, D090096A, D090096C, D090097C, D090097D, D090097E, D090097G, D090105B, D090105C, D090105D, D100007B, D100007C, D100008A, D100008B, D100009A, D100011A, D100011B, D100011C, D100016A, D100016B, D100016C, D100016D, D100022A, D100022B, D100022C, D100022D, D100026A, D100026B, D100026C, D100026D, D100026E, D100030A, D100030B, D100030C, D100035A, D100035B, D100035C, D100035D, D100043A, D100043B, D100043C, D100044A, D100044B, D100044C, D100044D, D100044E, D100045A, D100045B, D100045C, D100045D, D100059A, D100059B, D100059C, D100060A, D100060B, D100061A, D100061B, D100080A, D100081A, D100081AR, D100081AS, D110015A, D110015B, D110016A, D110045A, D110046A, D110047A, D110070A, D110071A, D110072A, D110090A, D110090B, D110091A, D110092A, D110108A, D110108B, D110108BR, D120015A, D120015B, D120027A, D120028A, D120029A, D120029B, D120038B, D120039A, D120056B, D120068A, D120068B, E090105A, E100016A, E100016B, E100035A, E100035B, E100035C, E100045A, E100045B, E100059A, E100059B, E100060A, E100060B, E100060C, E100061A, Amvisc 1.2% Sodium Hyaluronate ( Model 59051, 59081, 59051L, 59081L) Amvisc Plus 1.6% Sodium Hyaluronate (Model 60081, 60051, 60051L, 60081L)
Worldwide distribution: USA including DC and PR; and countries of: Canada and Mexico accounts not available at this time.
2,515,760