FDA Enforcement Class II Terminated

LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small bone fragments

Recall: Z-0669-2018 · Reported February 28, 2018

Enforcement

Recall Number
Z-0669-2018
Event ID
79021
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Synthes, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 28, 2018
Initiation Date
August 7, 2017
Classification Date
February 21, 2018
Termination Date
July 25, 2018
Address
1302 Wrights Ln E, N/A, West Chester, PA, 19380-3417, United States

Description

LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small bone fragments

Reason

May have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is drilled without the drill sleeve, the pilot hole could be off-center which could result in screw and/or plate loosening.

Code Info

Lot numbers FT00227 9972971 H028400 H161606 H288454

Distribution

Distribution to US, Canada, and Germany.

Quantity

567 units