Habib EUS RFA, 20mm electrode active tip, Material Number/UPN: 6700 - Product Usage: The Habib EUS RFA is indicated for coagulation and ablation of soft tissue when used in conjunction with compatible radiofrequency generator.
Enforcement
- Recall Number
- Z-0668-2020
- Event ID
- 84375
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Boston Scientific Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 18, 2019
- Initiation Date
- November 18, 2019
- Classification Date
- December 11, 2019
- Termination Date
- June 25, 2020
- Address
- 300 Boston Scientific Way, N/A, Marlborough, MA, 01752-1291, United States
Description
Habib EUS RFA, 20mm electrode active tip, Material Number/UPN: 6700 - Product Usage: The Habib EUS RFA is indicated for coagulation and ablation of soft tissue when used in conjunction with compatible radiofrequency generator.
During testing on the Habib EUS RFA device, a temperature increase was observed at the tip of the EUS-FNA needle. It was determined that capacitive 5 coupling presents a risk of energy transfer from the Habib EUS RFA active tip to the introducer device. This capacitive coupling could result in thermal injury to tissue outside of the targeted treatment area due to its use with an un-insulated EUS-FNA needle requiring little or no medical intervention.
Material Number/UPN: 6700 GTIN: 5060271140596 Batches: 17-C01 (Expiration: 10-Jul-2020), 17-C03 (Expiration: 12-Nov-2020), 17-C04 (Expiration: 18-Feb-2021), 17-C05 (Expiration: 18-Feb-2021), 17-C09 (Expiration: 19-Feb-2021)
Worldwide distribution - US Nationwide distribution in the states of AL, AZ, CA, CO, CT, FL, NJ, NY, MA, MD, MN, TN, TX and countries of Canada, India, Australia, Great Britain, Greece, Spain.