FDA Enforcement Class II Terminated

Habib EUS RFA, 10mm electrode active tip, Material Number/UPN: 6500 - Product Usage: The Habib EUS RFA is indicated for coagulation and ablation of soft tissue when used in conjunction with compatible radiofrequency generator.

Recall: Z-0667-2020 · Reported December 18, 2019

Enforcement

Recall Number
Z-0667-2020
Event ID
84375
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Boston Scientific Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 18, 2019
Initiation Date
November 18, 2019
Classification Date
December 11, 2019
Termination Date
June 25, 2020
Address
300 Boston Scientific Way, N/A, Marlborough, MA, 01752-1291, United States

Description

Habib EUS RFA, 10mm electrode active tip, Material Number/UPN: 6500 - Product Usage: The Habib EUS RFA is indicated for coagulation and ablation of soft tissue when used in conjunction with compatible radiofrequency generator.

Reason

During testing on the Habib EUS RFA device, a temperature increase was observed at the tip of the EUS-FNA needle. It was determined that capacitive 5 coupling presents a risk of energy transfer from the Habib EUS RFA active tip to the introducer device. This capacitive coupling could result in thermal injury to tissue outside of the targeted treatment area due to its use with an un-insulated EUS-FNA needle requiring little or no medical intervention.

Code Info

Material Number/UPN: 6500 GTIN: 5060271140725 Batches: 17-C02 (Expiration: 9-Jul-2020), 17-C06 (Expiration: 19-Feb-2021), 17-C07 (Expiration: 19-Feb-2021), 17-C08 (Expiration: 20-Feb-2021)

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AZ, CA, CO, CT, FL, NJ, NY, MA, MD, MN, TN, TX and countries of Canada, India, Australia, Great Britain, Greece, Spain.

Quantity

139