FDA Enforcement Class II Terminated

Alinity i Total Beta-hCG Reagent Kit, List Number 07P5121

Recall: Z-0666-2019 · Reported January 2, 2019

Enforcement

Recall Number
Z-0666-2019
Event ID
81632
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Abbott Ireland Diagnostics Division
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 2, 2019
Initiation Date
November 21, 2018
Classification Date
December 22, 2018
Termination Date
April 1, 2021
Address
Lisnamuck, Co. Longford, N/A, Ireland

Description

Alinity i Total Beta-hCG Reagent Kit, List Number 07P5121

Reason

Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.

Code Info

UDI/Lot Numbers (01)00380740160616 (17)190807(10)92056UI00 / 92056UI00

Distribution

AR, FL, GA, MD, OK, VA

Quantity

27 units