FDA Enforcement Class II Ongoing

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Recall: Z-0664-2025 · Reported December 18, 2024

Enforcement

Recall Number
Z-0664-2025
Event ID
95603
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Micro-X Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 18, 2024
Initiation Date
May 31, 2023
Classification Date
December 12, 2024
Address
Unit 14 6 Mab Eastern Promenade, Tonsley, N/A, Australia

Description

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Reason

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Code Info

UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿

Distribution

Worldwide and US Nationwide Distribution: NJ, PR, CA

Quantity

5