Philips Respironics V60 Ventilator Part Number 1053617
Enforcement
- Recall Number
- Z-0662-2022
- Event ID
- 89490
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Respironics California, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 9, 2022
- Initiation Date
- January 24, 2022
- Classification Date
- February 25, 2022
- Address
- 2271 Cosmos Ct, Carlsbad, CA, 92011-1517, United States
Description
Philips Respironics V60 Ventilator Part Number 1053617
A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient. This failure may either activate both visual and audible alarms, or it may not sound or display an alarm (a silent shutdown).
The Component of the Ventilators that is impacted is: N 1034132 BRACKET, CAPACITOR,MOTOR DRIVER PCB Model V60 Part Number: 1053617 Serial Numbers/UDI Code: MX0016593 / (01)00884838020054(21)MX0016593 MX0016797 / (01)00884838020054(21)MX0016797 MX0016824 / (01)00884838020054(21)MX0016824 MX0016827 / (01)00884838020054(21)MX0016827 MX0016830 / (01)00884838020054(21)MX0016830 MX0016839 / (01)00884838020054(21)MX0016839 MX0016840 / (01)00884838020054(21)MX0016840 MX0016841 / (01)00884838020054(21)MX0016841 MX0016845 (01)00884838020054(21)MX0016845 MX0016853 (01)00884838020054(21)MX0016853 MX0016867 (01)00884838020054(21)MX0016867 MX0016870 (01)00884838020054(21)MX0016870 MX0016871 (01)00884838020054(21)MX0016871 MX0016872 (01)00884838020054(21)MX0016872 MX0016875 (01)00884838020054(21)MX0016875 MX0016878 (01)00884838020054(21)MX0016878
U.S.: AR, CA, FL, GA, IN, LA, MD, MO, NY, PA, SC, TN and TX. O.U.S. N/A
16 systems