FDA Enforcement Class II Terminated

Fuji Computed Radiography Mammography Suite, FCRMS (for CR-IR363AWS) The Fuji Computed Radiography Mammography Suite (FCRMS) is a software device that, in conjunction with a specified Fuji Computed Radiography system forms the Fuji Computed Radiography for mammography (FCRm) device. FCRm with a dedicated mammographic x-ray machine generates digital mammographic images that can be used for screening and diagnosis of breast cancer. It is intended for use in the same clinical applications as traditional screen-film based mammographic (SFM) systems. The mammographic images can be interpreted by a qualified physician using either hardcopy film or softcopy display at a workstation.

Recall: Z-0662-2018 · Reported February 28, 2018

Enforcement

Recall Number
Z-0662-2018
Event ID
79078
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Fujifilm Medical Systems U.S.A., Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 28, 2018
Initiation Date
August 4, 2017
Classification Date
February 20, 2018
Termination Date
April 1, 2019
Address
419 West Ave, N/A, Stamford, CT, 06902-6343, United States

Description

Fuji Computed Radiography Mammography Suite, FCRMS (for CR-IR363AWS) The Fuji Computed Radiography Mammography Suite (FCRMS) is a software device that, in conjunction with a specified Fuji Computed Radiography system forms the Fuji Computed Radiography for mammography (FCRm) device. FCRm with a dedicated mammographic x-ray machine generates digital mammographic images that can be used for screening and diagnosis of breast cancer. It is intended for use in the same clinical applications as traditional screen-film based mammographic (SFM) systems. The mammographic images can be interpreted by a qualified physician using either hardcopy film or softcopy display at a workstation.

Reason

FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.

Code Info

363AWS. Released Software Versions: V5.1, V6.0. Not Released Software Versions: V5.0, V5.2, V6.1, V7.0.

Distribution

Worldwide Distribution.

Quantity

71 units (68 units Domestic. 3 units Foreign) in total