FDA Enforcement
Class II
Terminated
IPELA FHD PTZ TILE KITS with Axis cameras used in combination with Natus Neuroworks/Sleepworks software
Recall: Z-0661-2022
·
Reported March 2, 2022
Enforcement
- Recall Number
- Z-0661-2022
- Event ID
- 89425
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 2, 2022
- Initiation Date
- December 2, 2021
- Classification Date
- February 22, 2022
- Termination Date
- September 18, 2024
- Address
- 2568 Bristol Cir, Oakville, N/A, Canada
Description
IPELA FHD PTZ TILE KITS with Axis cameras used in combination with Natus Neuroworks/Sleepworks software
Reason
Tile kits that allow one to install and situate a video camera within a patient environment are not fitting properly in some domestic institution ceiling tile grids that have a thinner style grid frame. The ceiling tile may suddenly fall from the ceiling and cause a safety risk to the patient or end user.
Code Info
All serial numbers
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, IL, TX, MI, MA, NE, OK, WV, MO, VA, CT, IN, HI, PA, CA, MD, OH, CO, SC, PR, NV, NJ, OR, LA, DC, NC and the countries of Nepal, Viet Nam, Canada, Saudi Arabia, Indonesia, Germany, Korea, Republic of
Quantity
221