ASPIRE Cristalle (for FDR-3000AWS) The Fujifilm Digital Mammography System, ASPIRE Cristalle (FDR MS-3500) generates full field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.
Enforcement
- Recall Number
- Z-0661-2018
- Event ID
- 79078
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Fujifilm Medical Systems U.S.A., Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 28, 2018
- Initiation Date
- August 4, 2017
- Classification Date
- February 20, 2018
- Termination Date
- April 1, 2019
- Address
- 419 West Ave, N/A, Stamford, CT, 06902-6343, United States
Description
ASPIRE Cristalle (for FDR-3000AWS) The Fujifilm Digital Mammography System, ASPIRE Cristalle (FDR MS-3500) generates full field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.
FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.
3000AWS. Released Software Versions: V5.1, V5.2, V6.0, V6.1, V7.0. Not Released Software Versions: V5.0.
Worldwide Distribution.
71 units (68 units Domestic. 3 units Foreign) in total