FDA Enforcement
Class I
Terminated
5000 Series Freedom Drivers. Freedom Driver System for Temporary Total Artificial Heart (TAH-t). Part number 595000-001
Recall: Z-0659-2017
·
Reported December 14, 2016
Enforcement
- Recall Number
- Z-0659-2017
- Event ID
- 75600
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- SynCardia Systems Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Telephone
- Report Date
- December 14, 2016
- Initiation Date
- October 21, 2016
- Classification Date
- December 8, 2016
- Termination Date
- February 21, 2017
- Address
- 1992 E Silverlake Rd, N/A, Tucson, AZ, 85713-3865, United States
Description
5000 Series Freedom Drivers. Freedom Driver System for Temporary Total Artificial Heart (TAH-t). Part number 595000-001
Reason
The Main Printed Circuit Board Assembly (PCBA) of the affected Freedom Drivers may fail and cause the Freedom Driver to stop functioning without visual or audible alarms, resulting in the loss of life-sustaining function.
Code Info
S/N 5002 UDI (01)00858000003121(21)5002 S/N 5005 UDI (01)00858000003121(21)5005 S/N 5008 UDI (01)00858000003121(21)5008 S/N 5009 UDI (01)00858000003121(21)5009 S/N 5010 UDI (01)00858000003121(21)5010
Distribution
US distribution to Virginia and Arizona.
Quantity
5