FDA Enforcement Class II Terminated

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Recall: Z-0658-2022 · Reported March 2, 2022

Enforcement

Recall Number
Z-0658-2022
Event ID
89465
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Nobel Biocare Usa Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 2, 2022
Initiation Date
December 17, 2021
Classification Date
February 18, 2022
Termination Date
October 25, 2024
Address
22715 Savi Ranch Pkwy, Yorba Linda, CA, 92887-4609, United States

Description

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Reason

Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.

Code Info

Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497

Distribution

Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MN, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI. The countries of Canada, Mexico and Turkey.

Quantity

702 pieces