FDA Enforcement Class II Terminated

PKG, GRASPER FORCEPS, SERRATED, P/N 0250080286. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Recall: Z-0656-2015 · Reported December 24, 2014

Enforcement

Recall Number
Z-0656-2015
Event ID
69838
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Endoscopy
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 24, 2014
Initiation Date
November 17, 2014
Classification Date
December 17, 2014
Termination Date
July 17, 2015
Address
5900 Optical Ct, N/A, San Jose, CA, 95138-1400, United States

Description

PKG, GRASPER FORCEPS, SERRATED, P/N 0250080286. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Code Info

246416 340639 344735 542132 544416 742144 841812 842091 845858 847684 940079 942282 943532 944666 944831 945443 1040391 1040415 1041478 1042250 1042663 1042664 1042665 1042669 1042677 1143123 1144211 1145077 1145684 1146119 1147041 1240547 1241427 1242129 1242645 1243802 1243954 1244189 1245024 1245604 1246054 1246180 1343405 1343556 1344028 1344292 1344459 1345527 1346103 1346649 1440443 1441169 1442254 1442805 645372

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Quantity

180,573 units