FDA Enforcement Class II Terminated

COBRA Fusion 50 Ablation System, COBRA Fusion 150 Ablation System, and COBRA Fusion Magnetic Retriever System. Model Numbers: 700-001, 700-001S, 700-001MI, 700-002, and 700-003. Intended to ablate cardiac tissue during cardiac surgery using radiofrequency (RF) energy when connected directly to the Estech Electrosurgical unit (ESU).

Recall: Z-0653-2017 · Reported November 30, 2016

Enforcement

Recall Number
Z-0653-2017
Event ID
75373
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
AtriCure, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 30, 2016
Initiation Date
September 22, 2016
Classification Date
November 22, 2016
Termination Date
February 8, 2018
Address
7555 Innovation Way, N/A, Mason, OH, 45040-9695, United States

Description

COBRA Fusion 50 Ablation System, COBRA Fusion 150 Ablation System, and COBRA Fusion Magnetic Retriever System. Model Numbers: 700-001, 700-001S, 700-001MI, 700-002, and 700-003. Intended to ablate cardiac tissue during cardiac surgery using radiofrequency (RF) energy when connected directly to the Estech Electrosurgical unit (ESU).

Reason

Due to product design and process control, part of the device can separate and remain behind when the device is removed that can potentially cause permanent impairment or damage to body function or structure.

Code Info

Specifically all lots manufactured prior to February 2016. This information can be identified as the symbol for "date of manufacture" on the product label as: MM-YYYY, e.g. 02-2016 or earlier.

Distribution

Worldwide Distribution -- USA, in the following states AL, AR, AZ, CA, CO, CT, GA, FL, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA, WI, & VA; and, to the countries of Albania, Australia, Belgium, France, Germany, Ireland, Italy, Norway, Netherlands, Poland, Sweden, Turkey, United Arab Emirates, & United Kingdom.

Quantity

5,263