FDA Enforcement Class II Terminated

30 deg, 8mm Endoscope, IS4000: Model numbers: 470027-34, 470027-40, 470027-41. General and Plastic Surgery: The da Vinci Xi endoscope produces high definition (HD), visible light (VIS). The HD video is processed by the system electronics in the Vision Cart and displayed on the Surgeon Console 3D viewer and Vision Cart touchscreen. Endoscopes are available with either a 0 degree or 30 degree tips.

Recall: Z-0653-2016 · Reported January 20, 2016

Enforcement

Recall Number
Z-0653-2016
Event ID
72776
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Intuitive Surgical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 20, 2016
Initiation Date
November 23, 2015
Classification Date
January 14, 2016
Termination Date
April 5, 2016
Address
1266 Kifer Rd Bldg 100, N/A, Sunnyvale, CA, 94086-5304, United States

Description

30 deg, 8mm Endoscope, IS4000: Model numbers: 470027-34, 470027-40, 470027-41. General and Plastic Surgery: The da Vinci Xi endoscope produces high definition (HD), visible light (VIS). The HD video is processed by the system electronics in the Vision Cart and displayed on the Surgeon Console 3D viewer and Vision Cart touchscreen. Endoscopes are available with either a 0 degree or 30 degree tips.

Reason

This correction to the da Vinci Xi I&A User Manual communicates proper use of the da Vinci Xi 8 mm Endoscopes in combination with the da Vinci 8 mm and 12 mm cannulae, cannula reducers, and monopolar energy instruments. It informs customers of scenarios that may result in unintended patient harm due to capacitive coupling from monopolar energy sources.

Code Info

Model numbers: 470027-34, 470027-40, 470027-41. All lots.

Distribution

Worldwide Distribution - US (nationwide) and Internationally to Australia, Austria, Belgium, Cyprus, Czech Republic, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Japan, Monaco, Netherlands, Norway, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, and Venezuela. .

Quantity

2624 total all models