FDA Enforcement Class II Ongoing

Abiomed Kit, 14Fr Introducer, 13cm&25cm, Sterile Part of Pump Set (0048-0040, 0048-0045, 1000080)-Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3046

Recall: Z-0652-2024 · Reported January 17, 2024

Enforcement

Recall Number
Z-0652-2024
Event ID
93501
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Abiomed, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 17, 2024
Initiation Date
November 7, 2023
Classification Date
January 5, 2024
Address
24 Cherry Hill Dr, Danvers, MA, 01923-2575, United States

Description

Abiomed Kit, 14Fr Introducer, 13cm&25cm, Sterile Part of Pump Set (0048-0040, 0048-0045, 1000080)-Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3046

Reason

Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis

Code Info

UDI-DI: N/A

Distribution

Nationwide Foreign: Country AT AU BE BR CA CA CH CZ DE DK ES FR FR HK IN IT MX MY NL NO PA SG TW UK

Quantity

2551 units within set