Vantage Panoramic X-Ray System used to perform dental and orthodontic radiological imaging The Vantage Panoramic X-Ray System is a class II extra oral X-Ray system used to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw and oral structures. When the system is equipped with the cephalometric option, the system will also provide cephalometric radiographic examinations for use in orthodontic treatment planning and evalu
Enforcement
- Recall Number
- Z-0651-2016
- Event ID
- 70765
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Midmark Corp dba Progeny Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 3, 2016
- Initiation Date
- September 4, 2014
- Classification Date
- January 25, 2016
- Termination Date
- December 29, 2017
- Address
- 675 Heathrow Dr, N/A, Lincolnshire, IL, 60069-4206, United States
Description
Vantage Panoramic X-Ray System used to perform dental and orthodontic radiological imaging The Vantage Panoramic X-Ray System is a class II extra oral X-Ray system used to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw and oral structures. When the system is equipped with the cephalometric option, the system will also provide cephalometric radiographic examinations for use in orthodontic treatment planning and evalu
Failure of the primary collimator may result in unintentional movement of the device. The unintentional movement will not result in the generation of an error message.
Model Number: 60-A0001; ZA05002 , ZA05003 , ZA05004 , ZA05005 , ZA05006 , ZA05007 , ZA05008 , ZA05009 , ZA05010 , ZA05011 , ZA05012 , ZA05013 , ZA05014 , ZA05015 , ZA05016 , ZA05017 , ZA05018 , ZA05019 , ZA05020 , ZA05021 , ZA05022 , ZA05023 , ZB51001 , ZB51002 , ZB51003 , ZB51004 , ZB51007 , ZB51008 , ZB51009 , ZB51010 , ZB51011 , ZB51012 , ZB51013 , ZB51014 , ZB51015 , ZB51016 , ZB51017 , ZB51018 , ZB51019 , ZB51021 , ZB51023 , ZB51024 , ZB51025 , ZB51026 , ZB51027 , ZB51028 , ZB51029 , ZB51030 , ZB51031 , ZB51032 , ZB51034 , ZB51035 , ZB51036 , ZB52001 , ZB52002 , ZB52003 , ZB52004 , ZB52005 , ZB52006 , ZB52007 , ZB52008 , ZB52009 , ZB52010 , ZB52011 , ZB52012 , ZB52013 , ZB52014 , ZB52015 , ZB52016 , ZB52017 , ZB52018 , ZB52019 , ZB52020 , ZB52021 , ZB52022 , ZB52023 , ZB92002 , ZB92003 , ZB92004 , ZB92005 , ZB92006 , ZB92007 , ZB92008 , ZB92009 , ZB92011 , ZC38001 , ZC38002 , ZC38003 , ZC38004 , ZC38005 , ZC38007 , ZC38008 , ZC38009 , ZC38010 , ZC38011 , ZC38013 , ZC38015 , ZC38016 , ZC38017 , ZC38018 , ZC38019 , ZC38020 , ZC38021 , ZC38022 , ZC39001 , ZC39002 , ZE09001 , ZE09002 , ZE09003 , ZE09004 , ZE09006 , ZE09007 , ZE09008 , ZE09009 , ZE09010 , ZE09011 , ZE09012 , ZE09014 , ZE09015 , ZE09016 , ZE09017 , ZE09018 , ZE09019 , ZE09020 , ZE47001 , ZE47002 , ZE47003 , ZE47004 , ZE47005 , ZE47006 , ZE47007 , ZE47008 , ZE47009 , ZE47010 , ZE47011 , ZE47012 , ZE47013 , ZE47014 , ZE47015 , ZE47016 , ZE47017 , ZE47018 , ZE47019 , ZE47020 , ZE47021 , ZE77001 , ZE77002 , ZE77003 , ZE77004 , ZE77005 , ZE77006 , ZE77007 , ZE77008 , ZE77009 , ZE77010 , ZF41001 , ZF41002 , ZF41003 , ZF41004 , ZF41005 , ZF41006 , ZF41007 , ZF41008 , ZF41009 , ZF41010 , ZF41011 , ZF41012 , ZF41013 , ZF41014 , ZF41015 , ZF41016 , ZF41017 , ZF41018 , ZF41019 , ZF41020 , ZF41021 , ZF41022 , ZF41023 , ZF41024 , ZF95001 , ZF95002 , ZF95003 , ZF95004 , ZF95005 , ZF95006 , ZF95007 , ZF95008 , ZF95009 , ZF95010 , ZF95011 , ZF95012 , ZF95013 , ZF95014 , ZF95015 , ZF95016 , ZF95017 , ZG61001 , ZG61002 , ZG61003 , ZG61004 , ZG61005 , ZG61006 , ZG61007 , ZG61008 , ZG61009 , ZG61010 , ZG61011 , ZG61012 , ZG61013 , ZG61014 , ZG61015 , ZG61016 , ZG61017 , ZG61018 , ZG61019 , ZG61020 , ZG61021 , ZG61022 , ZG61023 , ZG61025 , ZG61026 , ZG61027 , ZG61028 , ZG61029 , ZG93001 , ZG93002 , ZG93003 , ZG93004 , ZG93005 , ZG93006 , ZG93008 , ZG93009 , ZG93010 , ZG93011 , ZG93012 , ZG93013 , ZG93014 , ZG93015 , ZG93016 , ZG93017 , ZG93018 , ZG93019 , ZG93020 , ZG93021 , ZG93022 , ZG93023 , ZG93024 , ZH20001 , ZH20002 , ZH20003 , ZH20004 , ZH20005 , ZH20006 , ZH20007 , ZH20008 , ZH20009 , ZH20010 , ZH20011 , ZH20012 , ZH20013 , ZH20014 , ZH20016 , ZH20017 , ZH20019 , ZH20020 , ZH20021 , ZH20022 , ZH20023 , ZH20024 , ZH20025 , ZH20026 , ZH20027 , ZH20028 , ZH20029 , ZH36001 , ZH36002 , ZH36003 , ZH36004 , ZH36005 , ZH36006 , ZH36007 , ZH36008 , ZH36009 , ZH36011 , ZH36012 , ZH36013 , ZH36014 , ZH36015 , ZH36016 , ZH36018 , ZH37002 , ZH37003 , ZH37004 , ZH37005 , ZH37006 , ZH37007 , ZH37008 , ZH37009 , ZH37010 , ZH37012 , ZH37013 , ZH37014 , ZH37015 , ZJ26001 , ZJ26002 , ZJ26003 , ZJ26004 , ZJ26005 , ZJ26006 , ZJ26007 , ZJ26008 , ZJ26009 , ZJ26010 , ZJ26011 , ZJ26012 , ZJ26013 , ZJ26014 , ZJ26015 , ZJ26016 , ZJ26017 , ZJ26018 , ZJ26019 , ZJ26020 , ZJ26021 , ZJ26022 , ZJ26023 , ZJ26024 , ZJ26025 , ZJ26026 , ZJ26027 , ZJ26028 , ZK04002 , ZK04003 , ZK04004 , ZK04005 , ZK04006 , ZK51001 , ZK51002 , ZK51003 , ZK51004 , ZK51005 , ZK51006 , ZK51007 , ZK51008 , ZK51009 , ZK51010 , ZK51011 , ZK51012 , ZK51013 , ZK51015 , ZK51016 , ZK51017 , ZK58001 , ZK58002 , ZK58003 , ZK58004 , ZK58005 , ZK58006 , ZK58007 , ZK58008 , ZK58009 , ZK58010 , ZK58011 , ZK58012 , ZK58013 , ZK58014 , ZK58015 , ZK58016 , ZK58017 , ZK58018 , ZK58019 , ZK58020 , ZK58021 , ZK58022 , ZK58023 , ZK58024 , ZM18018 , ZM81001 , ZM81002 , ZM81003 , ZM81004 , ZM81005 , ZM81006 , ZM81007 , ZM81008 , ZM81009 , ZM81010 , ZM81011 , ZM81012 , ZM81013 , ZM81014 , ZM81015 , ZM81016 , ZM81017 , ZM81018 , ZM81019 , ZM81020 , ZM81021 , ZM81022 , ZM81023 , ZM81025 , ZM81026 , ZM81027 , ZM81029 , ZM81030 , ZM81031 , ZN44001 , ZN44002 , ZN44003 , ZN44004 , ZN44005 , ZN44006 , ZN44007 , ZN44008 , ZN44009 , ZN44010 , ZN44011 , ZN44012 , ZN44013 , ZN44014 , ZN44015 , ZN44016 , ZN44017 , ZN44018 , ZN44019 , ZN44020 , ZN44021 , ZN44022 , ZN44023 , ZN44024 , ZP01002 , ZP01003 , ZP01004 , ZP01005 , ZP01006 , ZP01007 , ZP01008 , ZP01011 , ZP01012 , ZP01013 , ZP01014 , ZP01015 , ZP01016 , ZP01017 , ZP01018 , ZP01019 , ZP01020 , ZP01021 , ZP01022 , ZP01023 , ZP01024 , ZP01025 , ZP01026 , ZP01027 , ZP01028 , ZP01029 , ZP01030 , ZP01031 , ZP01032 , ZP01033 , ZP01034 , ZP01035 , ZP01036 , ZP01037 , ZP01039 , ZP01041 , ZP01042 , ZP01043 , ZP0I034 , ZP42001 , ZP42002 , ZP42003 , ZP42004 , ZP42005 , ZP42006 , ZP42007 , ZP42008 , ZP42009 , ZP42010 , ZP42011 , ZP42012 , ZP42014 , ZP42015 , ZP42016 , ZP42017 , ZP42018 , ZP42019 , ZP42020 , ZP42021 , ZP42022 , ZR45001 , ZR45002 , ZR45003 , ZR45004 , ZR45005 , ZR45006 , ZR45007 , ZR45008 , ZR45009 , ZR45010 , ZR45011 , ZR45012 , ZR45013 , ZR45014 , ZR45015 , ZR45016 , ZR45017 , ZR45018 , ZR45019 , ZR45020 , ZR45021 , ZR45022 , ZR45023 , ZR45024 , ZR45025 , ZR45026 , ZR48001 , ZR48002 , ZR48003 , ZR48004 , ZR48005 , ZR48006 , ZR48007 , ZR48008 , ZR48009 , ZR48010 , ZR48011 , ZR48012 , ZR48013 , ZR48014 , ZR48015 , ZR48016 , ZR48017 , ZR48018 , ZR48019 , ZR48020 , ZR48021 , ZR48022 , ZS43001 , ZS43002 , ZS43003 , ZS43004 , ZS43005 , ZS43006 , ZS43007 , ZS43008 , ZS43009 , ZS43010 , ZS43011 , ZS43012 , ZS43013 , ZS43014 , ZS43015 , ZS43016 , ZS43017 , ZS43018 , ZS43019 , ZS54002 , ZS54004 , ZS54005 , ZS54006 , ZS54007 , ZS54008 , ZS54009 , ZS54010 , ZS54011 , ZS54012 , ZS54013 , ZS54014 , ZS54015 , ZS54016 , ZS54017 , ZS54019 , ZS54020 , ZS54021 , ZS59001 , ZS59002 , ZS59003 , ZS59004 , ZS59005 , ZS59006 , ZS59007 , ZS59008 , ZS59009 , ZS59010 , ZS59011 , ZS59012 , ZS59013 , ZS59014 , ZS59015 , ZS59016 , ZS59017 , ZS59018 , ZS59019 , ZT18001 , ZT18002 , ZT18003 , ZT18004 , ZT18005 , ZT18006 , ZT18007 , ZT18008 , ZT18009 , ZT18010 , ZT18011 , ZT18012 , ZT18013 , ZT18014 , ZT18015 , ZT18016 , ZT18017 , ZT18018 , ZT18019 , ZT18020 , ZT18021 , ZT18022 , ZT18023 , ZT18024 , ZT18025 , ZT18026 , ZT18027 , ZT18028 , ZT18029 , ZT18030 , ZT18031 , ZT18032 , ZT18033 , ZT18034 , ZT18035 , ZT18036 , ZT18037 , ZT18038 , ZT18039 , ZT18040 , ZT27001 , ZT27002 , ZT27003 , ZT27004 , ZT27005 , ZT27006 , ZT27007 , ZT27008 , ZT27009 , ZT27010 , ZT27011 , ZT27012 , ZT27013 , ZT27014 , ZT27015 , ZT27016 , ZT27017 , ZT27018 , ZT27019 , ZT27020 , ZT27021 , ZT27022 , ZT27023 , ZT27024 , ZU33001 , ZU33002 , ZU33003 , ZU33004 , ZU33005 , ZU33007 , ZU33008 , ZU33009 , ZU33011 , ZU33012 , ZU33013 , ZU33014 , ZU33015 , ZU33016 , ZU33017 , ZU33018 , ZU33019 , ZU33020 , ZU33021 , ZU33022 , ZU33023 , ZU33024 , ZU33025 , ZU33026 , ZU33027 , ZU33029 , ZU33030 , ZU33031 , ZU33032 , ZU33033 , ZU33034 , ZU33035 , ZV29001 , ZV29002 , ZV29003 , ZV29004 , ZV29005 , ZV29006 , ZV29007 , ZV29008 , ZV29009 , ZV29010 , ZV78001 , ZV78002 , ZV78003 , ZV78004 , ZV78006 , ZV78007 , ZV78008 , ZV78009 , ZV78010 , ZV78011 , ZV78012 , ZV78013 , ZV78014 , ZV78015 , ZV78016 , ZV78017 , ZV78018 , ZV78019 , ZV78020 , ZV78021 , ZV78022 , ZW15001 , ZW15002 , ZW15003 , ZW15004 , ZW15005 , ZW15006 , ZW15007 , ZW15008 , ZW15009 , ZW15011 , ZW15012 , ZW15013 , ZW15014 , ZW15015 , ZW15016 , ZW15017 , ZW15018 , ZW15019 , ZW15020 , ZW15021 , ZW15022 , ZW34002 , ZW34003 , ZW34004 , ZW34006 , ZW34007 , ZW34008 , ZW34009 , ZW34010 , ZW34011 , ZW34013 , ZW34015 , ZW34016 , ZW34017 , ZW34018 , ZW34019 , ZW34020 , ZW34021 , ZW34022 , ZW74007 , ZW74008 , ZW74009 , ZW74010 , ZW74011 , ZY49001 , ZY49002 , ZY49003 , ZY49004 , ZY49005 , ZY49006 , ZY49007 , ZY49009 , ZY49011 , ZY49012 , ZY49014 , ZY49016 , ZY49017 , ZY49018 , ZY49019 , ZY49020 , ZY49021 , ZZ02001 , ZZ02002 , ZZ02003 , ZZ02004 , ZZ02005 , ZZ02006 , ZZ02007 , ZZ02008 , ZZ02009 , ZZ02011 , ZZ02012 , ZZ02013 , ZZ02014 , ZZ02015 , ZZ02016 , ZZ02017 , ZZ02018 , ZZ02019 , ZZ02020 , ZZ02021 , ZZ02022 , ZZ02023 , ZZ02024 , ZZ08001 , ZZ08002 , ZZ08003 , ZZ08004 , ZZ08005 , ZZ08006 , ZZ08007 , ZZ08008 , ZZ08009 , ZZ08010 , ZZ08011 , ZZ08012 , ZZ08013 , ZZ08014 , ZZ08015 , ZZ08016 , ZZ08018 , ZZ08019 , ZZ08020 , ZZ08021 , ZZ08022 , ZZ08023 , ZZ69000 , ZZ69003 , ZZ69004 , ZZ69005
Worldwide Distribution - US including AL, AR, AZ, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI; and Internationally to Australia, Belgium, Canada, Czech Republic, France, Hungary, Lithuania, New Zealand, Poland, Russia, Saudi Arabia, South Africa, Switzerland, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
832 systems