FDA Enforcement Class II Terminated

Catalys Precision Laser System (United States) - Catalys-U: Catalys Precision Laser System (International) - Catalys I 200-240V ~, 50/60 Hz, 15A, (2) 15A Resettable CB: Tested and complies with FCC Part 15 Class A. OptiMedica 1310 Moffett Park Drive Sunnyvale, CA 94089 The OptiMedica Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens.

Recall: Z-0649-2014 · Reported January 15, 2014

Enforcement

Recall Number
Z-0649-2014
Event ID
67150
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Optimedica Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 15, 2014
Initiation Date
December 20, 2013
Classification Date
January 8, 2014
Termination Date
June 26, 2014
Address
1310 Moffett Park Dr, Sunnyvale, CA, 94089-1133, United States

Description

Catalys Precision Laser System (United States) - Catalys-U: Catalys Precision Laser System (International) - Catalys I 200-240V ~, 50/60 Hz, 15A, (2) 15A Resettable CB: Tested and complies with FCC Part 15 Class A. OptiMedica 1310 Moffett Park Drive Sunnyvale, CA 94089 The OptiMedica Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens.

Reason

Potential for unexpected heating of the transformer mounting bolt located within the system chassis and the subsequent melting of a non-slip computer shelf mat, which may manifest an odor and visible smoke.

Code Info

All units shipped prior to Octotber 18, 2013. US Serial numbers: 44055311, 44055611, 44055811, 44056012, 44056312, 44056612, 44056712, 44056812, 44057312, 44057812, 44058012, 44058212, 44058312, 44059212, 44059312, 44059612, 44059712, 44059812, 44059912, 44060012, 44060112, 44060312, 44060412. International Serial numbers: 44000811, 44055211, 44055411, 44055711, 44055911, 44056112, 44056212, 44056412, 44056512, 44056912, 44057012, 44057112, 44057212, 44057412, 44057512, 44057612, 44057712, 44057912, 44058112, 44058512, 44058612, 44058912, 44059012. 44059112 44059412 44059512 44060212 44060512 44061112 44061212 44061312 44061512 44061712 44061913 44062113 44062313 44062513 44062913 44063013 44060612 44060712 44060912 44061012 44061412 44061612 44061812 44062013 44062213 44062613 44062813 44063313 44063513 44063613 44063713 44063913.

Distribution

Worldwide Distribution: US (nationwide) including states of: TX, NC, OH, IA, UT, KY, ME, CA, NY, MA. NJ, IL, PA,, TN, FL, MO, GA, HI, SC, NE, MI, and GA and countries of: Dominican Republic, Germany, Switzerland, Australia, Spain, Japan, United Kingdom, Canada, France, The Netherlands, Austria, and Singapore.

Quantity

78 (39 in US, 39 internationally)